![Vernacare Vortex+ Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/vernacare/vortex/vortex_technical-manual_1006377023.webp)
Page | 23
SECTION 9: Hopper Vent Arrangement
1.1
2
1.2
1.3
1.4
1.5
1.6
KIT
REF
PART DESCRIPTION
KIT NUMBER
ITEM NUMBER
1
Vent Connector Kit
9207007
1.1
Non Return Valve
9757030
1.2
Vent Connector
9200334
1.3
‘O’ Ring
9752018
1.4
Vent Deflector
9200325
1.5
M4 x 8 Lg Slotted Pan Head Screw (x2)
9750064
1.6
M4 Nylon Washer x2
9752022
2
2
16mm PVC Tubing 1.5m
9200721
Detail of Vent Connection
Situated at the front top right hand side of the hopper is a vent connector (2) which allows
the foul air which is displaced whilst water is filling the hopper, to be vented into the
drain.
The vent connector is fitted with a non - return valve (push fit into connector with the
arrow on its body pointing into the connector). This valve prevents air from being
expelled to atmosphere during water filling but allows air to be drawn into the hopper
during the drain cycle when a vacuum would otherwise be created.
The connector is secured to the hopper via two M4 screws and two nylon washers, to
prevent leakage an O-ring is used to seal the connector to the hopper and is located in
a groove on the inner face of the connector