24
10 Tub Vent Arrangement
Figure 8 – Vent Assembly
Detail of Vent Connection
Situated at the front top right hand side of the hopper is a vent connector (2) which allows
the foul air which is displaced whilst water is filling the hopper, to be vented into the drain.
The vent connector is fitted with a non - return valve (1) (push fit into connector with the
arrow on its body pointing into the connector). This valve prevents air from being expelled
to atmosphere during water filling but allows air to be drawn into the hopper during the
drain cycle when a vacuum would otherwise be created.
The connector is secured to the hopper via two M4 screws (5) and two nylon washers (4),
to prevent leakage an O-ring (3) is used to seal the connector to the hopper and is
located in a groove on the inner face of the connector.
KIT
REF
PART DESCRIPTION
KIT NUMBER
ITEM NUMBER*
1
Vent Connector Kit
9207007
1.1
Non Return Valve
9757030
1.2
Vent Connector
9200334
1.3
‘O’ Ring
9752018
1.4
Vent Deflector
9200325
1.5
M4 x8 LG Slotted Pan Screw x2
9750064
1.6
M4 Nylon Washer x2
9752022
2
2
16mm PVC Tubing
9200721