13
2.7 Display screen icons and warning messages – explanations
RUNNING
The machine has been started and is running through its cycle.
END
The machine has finished the cycle and is ready to open.
BLOCKED BLADES
The motor invertor has defaulted to overload due to a blockage
of the machine cutter blades.
BLOCKED DRAIN
The pressure transducer has sensed an over pressure in the
chamber due to a blockage of the drain outlet.
LOW WATER
There is insufficient water in the water supply tank.
POWER FAIL
There has been an interruption to the mains power supply/cycle.
The motor is running and the cutter blades are rotating.
The water pump is running.
The deodoriser reservoir is empty.
The outlet diaphragm is open.
The outlet diaphragm is closed.
The water tank reservoir level is low.
The water tank reservoir is full.
The lid is open.
The lid is closed.
The drain is blocked.
The foot sensor has been operated.
(For details of the fault finding guide see section 14 page 29).