![Vernacare COMPACT+ Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/vernacare/compact/compact_technical-manual_1006373008.webp)
Vernacare Compact Plus Technical Manual-
Revision 2-Issued by N.Shields 14.2.2020
Page | 8
2.4.
Motor
0.55 KW; 3 Phase; 220-240v 50Hz/ 380-420v 50Hz; IP55; D71 frame size. Rating at
peak load 2.6 amps (nominal).
2.5. Inverter
AC inverter with either 110v or 230v supply voltage, 230v motor output and utilising
current monitoring.
2.6.
Pump
0.18 kW; 24vdc; 50 cycles. Protected by a 10 amp fuse.
2.7. Water requirements
Inlet supply: Cold water
ONLY
with flow rate of 8 l/min minimum. Nominal overall usage;
18 litres (+/- 10%) per (standard) cycle.
¾” BSP connections (male)
2.8.
Drainage
A 2
” inch/ 50mm drain pipe to be connected either directly, or via a swept bend, to a
standard
2”in/50mm drain pipe with a 1 in 25 or suitable fall into a soil drain.