![Vernacare COMPACT+ Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/vernacare/compact/compact_technical-manual_1006373038.webp)
Vernacare Compact Plus Technical Manual-
Revision 2-Issued by N.Shields 14.2.2020
Page | 38
11.2.
Setting ‘AUTO START’ & Adjusting Cycle Modes
SET TIME&DATE
CYCLE SETTINGS
COMMS SETTINGS
PRIME PUMPS
SERIAL NUMBER
SAVE&EXIT
AUTO START
SERVICE RESET
CYCLE SELECT
AUTO START
(ON)
/OFF
AUTO START
ON/
(OFF)
SERVICE RESET
(ON)
/OFF
SERVICE RESET
ON/(OFF)
STANDARD CYCLE
PLUS CYCLE
ECO CYCLE
SETTING CHANGED
BACK
Once you are happy with
your changes, scroll to
SAVE&EXIT then press
start to bring you out of
engineers mode.
Note that the alternate cycle modes are as follows:
Standard Cycle- Normal operation. 4 products/cycle.
Heavy Duty Cycle- Used only for busy wards when the machine is running continuously. 4
products/cycle.
Eco Mode- Used for light duty machines. 1 product only/cycle.
Please take care to select the appropriate cycle to ensure the machine operated both as safely and
economically as possible.
The service reset on/off function allows an engineer to rest the service due message which appears
every 5000 cycles.