MANUAL OF INSTRUCTIONS
8
VERLINDE reserves the right to alter or amend the above information without notice
09/2021
MESSAGES OF WARNING
[bAtEr]
The load level of the battery is very low.
Use the loader delivered for its loading.
When the level reaches a minimum value, the device
realises an automatic disconnection in order not to damage
the battery.
[CELUL]
During the start of the device, the weight detected superates
with 100% the maximum capacity allowed or there is a failure
in the weighing system.
[ALto ]
The gross weight is higer than the maximum capacity of the
device.
[BAJo ]
The negative gross weight is higher that the maximum
capacity of the device.
[SobrE]
Failure in the weighing system.
[ooooo]
Failure in the weighing system.
[Error]
1) One has tried to realise a sum of weight without
previously having unloaded the load.
2) During the adjustment of weight the sample introduced
is “0” or of a value higher than the capacity of the
device.
[ErrdV]
The characteristics of the cell or the adjustment do not adapt
with the value of the introduced sample weight.
Change the capacity and fraction of the device in order to obtain
less points of resolution and repeat the adjustment.
RELOAD OF BATTERIES
The system of feeding named Service 24 hours, is composed of 2 reloadable
interchangeable bat an exterior loader. The system allows to work with a battery while
uploading the other.
If one wants to increase the autonomy of work it can be defined in the paragraph of
configuration that operates the Automatic Disconnection after a period of having done the
weighing.
When the screen warns with the flashing message
"bAtEr"
, it indicates that the battery
is low and has to be substituted by the other one that is uploaded.
The extracted battery has to be connected with the exterior uploader.
The uploader has to be connected with a plug of 220 V c.a. during 8...12h.
THE LIFE OF THE BATTERIES DIMINISHES IN THE FOLLOWING CASES:
- When uploads are superior to 12 hours.