EMG Session Statistics
The EMG session statistics are used for assessment of muscular activities.
Start the session, follow the on-screen prompts, by the end of the session the
following statistics will be displayed. Press SET to scroll down the list of
statistics.
SE and T are displayed only when the unit is locked.
SE
- the number of completed sessions when the unit
was locked. If no sessions were completed and you
unlock the unit, SE will read zero and it will be the only
statistics screen you would see, as nothing else to be
presented.
T
- number of trials set for a session.
Work Average [µV]
- the overall average microvolts
achieved during all work periods of the session.
Generally the higher the Work average is, the better the
muscle performance.
Assessment: the good progress is when Work average
is getting higher and higher each day.
20
SE
T
5
WAV
67.0uv
RAV
0.8uv
Rest Average [µV]
- the overall average microvolts
during all rest periods of the session.
Generally the lower the Rest average is, the better the
muscle performance. It is very important how low you
can relax your muscles in terms of microvolts. Below
4µV a muscle is beginning to rest. If the Rest average is
above 4µV make sure you use the EMG reference lead
wire! The reading above 4µV commonly means the
muscle is overstimulated or tired after a longer EMG
training session.
Assessment: a good progress is when the Rest average
is getting lower and lower each day.
Microvolts
PROMPTS
inicial rest
“WRK”
“RST”
“RST”
“WRK”
WORK.1
REST.1
WORK.2
REST.2
release.1
onset.1
release.2
onset.2
[2 x Trials of: 4 sec Work and 4 sec Rest]
Time
[sec]