NeuroTrac
®
Simplex - Operation Manual
15
EMG Session settings
EMG
parameters
Text on
LCD
Available settings
Sound Volume
VOL
Value range: 0 - sound off, 1,2,3,4,5 - quiet, 5,6,7,8,9,10
- loud.
Typical value: 4.
This is the sound volume of beeps and prompts: Work,
Rest, Lock, PC-CONNECT, Erase
Work Time
WRK
Value range: [sec]: 2,3,4, ..., 99
Typical value: 5.
This is the time during which the patient is prompted
to keep the measured muscles contracted above the
threshold level.
Rest Time
RST
Value range: [sec]: 2,3,4, ..., 99.
Typical value: 5.
This is the time during which the patient is prompted to
keep the measured muscles relaxed below the threshold
level, ideally below 1uV.
Number of
Trials
TRS
Value range: 2,3,.., 99.
Typical value: 5.
This parameter defines how many repetitions of Work/
Rest cycles your treatment requires.
Audio
Biofeedback
type
FAB
FBL
FCO
FOF
Select from the following:
FAB - the EMG beeps will be audiable only above the
threshold.
FBL - the EMG beeps will be audiable only below the
threshold.
FCO - the EMG beeps will be audiable above and
below the threshold
FOF - no EMG beeps. Typical setting: FAB.
PC Data
output
DON
DOF
Select from the following:
DON: - switch ON to enable PC Software connection.
DOF: - switch OFF to save batteries.
Typical setting: DON.
EMG Notch
Filter
WDE
NRW
Select from the following:
WDE
(typical setting)
- wide filter - your EMG reading will
be more sensitive but it will also pick up heart beats.
NRW - narrow filter - your EMG reading will be filtered
to reduce the heart beat spikes on the EMG reading. Use
narrow filter if your electrodes are in close proximity to
heart (on chest, back, arms, vaginally or rectally).
To adjust the session parameters, press and hold SET button for 3 seconds.
The first parameter will be VOL (sound volume). To adjust the value, press
+/- buttons. To go to the next parameter in the table below, press SET button.
To save new settings, simply press ESC or OFF. The new settings will be
immediately stored.
Содержание NeuroTrac Simplex