background image

SPECIFICATIONS 

Model number 

Capacity 

Division 

T

 302 

   300g 

0.01g 

T

 -602 

   600g 

0.01g 

T

 -1002 

   1000g 

0.01g 

T

 -2001 

   2000g 

0.1g 

T

 -3001 

   3000g 

0.1g 

T

 -5201 

   5200g 

0.1g 

Platter 136x140mm 

Net/gross weight 

3.5lb/4.5lb 

 

Package 

Standard carton: 33.5

×

23

×

14(cm

3

6 Units in one box: 

48.5

×

39.5

×

46(cm

3

)

 

Operating Temperature 

0-40

(32-104

Power source 

6XAAor AC/DC 

Adapter 12V/500mA 

 
ERROR CODE 

 

 

COMPACT SCALE

/

BALANCE 

 

INSTRUCTION MANUAL 

 
 
 
 

 

 

 

Error 

Code 

Reason Trouble 

Shooting 

Err-O Overload 

Use 

within 

capacity. 

Err-Z 

1).Exceeded the zero 
tracking range when 
powering on. 
2).Forgot to install pan 
when performing 
linearity calibration. 

1). Pan should be empty on power up. 
Remove any weight from pan. 
2). Install pan and then recalibrate scale

linearity calibration

Err-S 

Inner code unstable 
when powering on. 

1) Make sure the scale is stable when 
powering on. 
2) Low battery, change or charge battery. 
3) Allow for longer warm up period. 

 

Err-L 

Calibrating without a 
weight or the weight   
is not the correct value. 

Place a calibration weight of the value 
shown on the display and then press 
calibration button. 

Err-P 

Counting setting error. 

Place the correct size parts sample when 
setting up counting mode. 

Содержание T-302

Страница 1: ...y Err Z 1 Exceeded the zero tracking range when powering on 2 Forgot to install pan when performing linearity calibration 1 Pan should be empty on power up Remove any weight from pan 2 Install pan and then recalibrate scale linearity calibration Err S Inner code unstable when powering on 1 Make sure the scale is stable when powering on 2 Low battery change or charge battery 3 Allow for longer warm...

Страница 2: ...TION When to calibrate 1 Press ON OFF to turn the power on for 15 minutes for warm up 2 With scale turned off press ON OFF to turn the power on 3 Display will show flashing digits and then 4 Press ZERO the display will show CAL 5 Press ZERO again the display will show X0000 g where X is the flashing digit 6 Press UNIT key to move the flashing digit to the right 7 Press PCS key to increase the flas...

Страница 3: ...an press PCS key to select zero range when turn on the scale press UNIT key to select percent PZR 0 2 3 4 5 10 20 100 FS Press ZERO key to set zero range press UNIT key to select percent KZR 3 4 5 10 20 Press ZERO key to set tare range press UNIT key select percent KTR 50 FS or KTR 100 FS Press ZERO key to enter overload protection setting press UNIT key to select overload protection value OVR 0 2...

Страница 4: ... please allow the balance to warm up for 10 minutes after first turning to power on so that the balance will function properly and accurately Calibration may be required before weighing Read CALIBRATION first and if necessary calibrate your scale balance for accurate weighing Weighing procedures 1 Press ON OFF to turn on the balance When power is turned on all display segments appear on LCD for a ...

Отзывы: