TROUBLESHOOTING - ERROR CODES
E 1
E 2
E 3
E 10
E 11
E 20
E 21
Eexx
The cuff is not secure/not
tight enough.
The cuff is wrapped too
tightly.
The cuff pressure is over-
inflated.
Excessive movement was
detected during measure-
ment.
An error occurred during
measurement and the re-
sults are inaccurate.
A calibration error has oc-
curred.
If an error code persists or is not remedied, please contact
Customer Care at 866-326-1313.
Re-apply the cuff and at-
tempt new measure-
ment.
Re-apply the cuff and at-
tempt new measure-
ment.
Relax for several min-
utes and attempt new
measurement
Remain still and attempt
new measurement.
Relax for several min-
utes and attempt new
measurement
A calibration error has
occurred.
ENGLISH •
25
CODE
CAUSE
SOLUTION