Rev 1.1 Cod. 00.009.126
32
ALARM
In the menu
Relay
, select by the cursor "Alarm" and press
ENTER
.
Select the desired operating mode of the relay with the ▼ ▲
buttons and press
ENTER
.
N.O = Normally Open (default value)
N.C = Normally Closed
If you want to esc programming and return to measurement,
press the ESC key for 3 seconds.
DELAY ON
In the menu
Relay
, select by the cursor "Delay On" and press
ENTER
.
Include or exclude this alarm by ▼▲ buttons and press
ENTER
.
If you want to esc programming and return to measurement,
press the ESC key for 3 seconds.
To INCLUDE OR EXCLUDE the other alarms follow the same steps of Delay On here above.
UR-OR
What is the UR-OR alarm?
It is possible to set an UR = Under Range threshold for the measure below which the display
will show the UR message.Similarly, a threshold of OR = Over Range can be set for the measure above which the display
will show the OR alarm message.
In both cases the relay will signal the alarm by closing or opening depending on the setting of the "Alarm" parameter. This
type of alarm does not influence the dosing of the pump.
In the menu
Relay
, select by the cursor "
UR-OR
" and press
ENTER
.
Include or exclude the alarm with the ▼ ▲ buttons and press
ENTER
.
If this alarm is included the display shows the programmed
values for UR and OR. If you want to change them, move the
cursor to select the parameter to be changed and press
ENTER
.
Use the ▼ ▲ keys to select the desired value digit by digit
and press
ENTER
each time.
If you want to esc programming and return to measurement,
press the
ESC
button for 3 seconds.