address value, access the level 2 list. In level 2 press the page key until the COMMS
parameter is displayed. Press up
down
to select the address value.
4.8.4
Alarm Option
When an alarm board is fitted, which consists of a relay with voltage free contacts, for
operator use, the contacts are taken to a panel plug on the control panel, wired as
indicated:
Key
C
Temperature Controller
F
Fuse (2A)
S
Supply
L
Load
*
Normally open relay
contacts
RO
Relay Output 240V 2A
MAX
The purpose of the 2 amp fuse is to break the circuit to prevent overloading on the
circuit due to high voltage.
The instrument configuration and parameters available to the operator depend on the
customer requirements.
4.9
Temperature Controller Replacement
Before handling the controller: wear an anti-static wrist strap or
otherwise avoid any possibility of damage to the unit by static electricity.
Refer to the detailed instructions supplied with the replacement
controller.
Ease apart the two lugs at the side; grip the instrument and withdraw it from its sleeve;
push in the replacement.
4.10
3216 Controller Navigation Diagram
The following diagram details how to navigate to the various menu options within the
3216 Controller. At each option, values can be set using the arrow keys.
25
4.0 3216 Controller
Содержание CARBOLITE GERO TZF 15/90/610
Страница 6: ...1 0 Symbols and Warnings Caution Double Pole Neutral Fusing 6 ...
Страница 50: ......
Страница 51: ...Notes Service Record Engineer Name Date Record of Work ...