8.6
Thermocouple Replacement
Disconnect the product from the power supply. Remove the appropriate panel
to gain access to the thermocouple connections. Make a note of the
thermocouple connections.
Thermocouple cable colour codings are:
thermocouple leg
colour
positive (type R)
orange
negative
white
Disconnect the thermocouple from its terminal block; retain any porcelain spacers.
Withdraw the thermocouple from its sheath and remove any broken bits of
thermocouple.
Re-assemble with a new thermocouple observing the colour coding.
Refit the element access panel.
8.7
Fuse Replacement
Fuses are marked on the wiring diagram with type codes, e.g. F1, F2. For more
information on fuses refer to section 11.0.
Depending on model and voltage, the different fuse types may or may not be fitted.
If any fuse has failed, it is advisable for an electrician to check the internal circuits.
Replace any failed fuses with the correct type. For safety reasons do not fit larger
capacity fuses without first consulting Carbolite Gero.
The fuses are located at the cable entry point. Remove the back panel or control box
back panel to gain access to the fuses.