![Vemco VR60 Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/vemco/vr60/vr60_hardware-manual_864184016.webp)
12
Setup
The setup routine contains the following features:
1. Set internal time-of-day clock
2. Turn on/off the printer, set the print interval and baud rate.
3. Setup transmitter parameters (frequency, coding type, units, slope, intercept) for up to twelve
transmitters. The transmitter parameters will remain in memory storage until changed.
4. Set logging interval. (Option 03B)
Enter the
Setup
mode by pressing the Shift key (red plus sign) and the SETUP key (top right corner
of keypad). The routine will then move through the available selections in the following order:
Time of Day, Printer, Log Interval (Option 03B), Blanking Interval, and Transmitters Setup. The
sections of SETUP pertaining to Option 03B is explained in the VR60 Memory section of this
manual.
Time of Day
The display shows the current time in the format days: hours: minutes: seconds (see example
below). The second line in the display shows the Y/N prompt. Select YES if the time shown is
correct, or select NO and follow the prompts to enter the correct time.
3: 8: 25: 16
TIME (Y/N)
The time of day should be reset each time the receiver is turned on if a printer is used, or if data
logging (Option 03B) is employed.
Printer
After the date has been selected, the printer status will be displayed as either:
No Printer (Y/N)
or
Printer (Y/N)
If there is no printer or storage device connected to the serial port, select YES at the “No Printer
(Y/N)” prompt. If a printed record of the data is desired, select NO at the “No Printer (Y/N)” and
YES at the “Printer (Y/N)” prompt. If a printer is selected, the unit will display the print interval
(time between printing data to the serial port) in seconds. The print interval may be changed by
selecting NO at the (Y/N) prompt and then entering the new interval (between 0 and 100 seconds).
NOTE:
When the print interval is zero all data is printed. For example, with a print interval of 10
seconds the data is only printed every 10 seconds.
With a printer selected, the unit will display the current baud rate. Press NEXT until the desired