AquaCare Twin: Installation
Preparing the AquaCare Twin For First Use
When training with the AquaCare Twin, powder could cause mess and damage to
sensitive equipment if the user is not accustomed to operating the machine. For first
use, either work near a sink or lay out a towel on the worksurface. Aim the handpiece
nozzle into the inlet of your evacuation unit. Always wear goggles when operating the
AquaCare Twin.
See INTRODUCTION: SAFETY WARNINGS, Pg 5.
A. Check that all of the INSTALLATION sections have been completed and the
INSTALLATION TEST has been successful.
B. Fit the left dosing chamber with a sodium bicarbonate powder cartridge (Prophy).
C. Fit the right dosing chamber with either 29 or 53 micron Aluminium Oxide powder
cartridge (Prep).
D. Switch the powder chamber selector to left (Prophy) chamber.
E. Make sure you are in full control of the handpiece.
F. Switch the On/Off selector to ON position.
G. Press the foot pedal fully down to the CUT position 3. You will hear the vibrator start
to operate. The nozzle will start to spray a mixture of air, fluid and powder.
H. Turn the Media volume dial, observing the spray. On a paper towel, you should see a
ring of white powder inside an outer ring of fluid. The Media volume range as shown
by the media gauge varies between:
Slightly under midway position - Low powder content, slow cutting/cleaning
Midway position - Medium powder content, cutting
Slightly above midway position - High powder content, very fast cutting
I. Turn the powder chamber selector to the right (Prep) chamber to check that the
other powder flows, then repeat step H to check the right media dial settings.
J. The powder needs to be purged whenever you replace a powder cartridge. The
fluid needs to be purged before first use and at the start of every day. See
MAINTENANCE: PURGING THE AQUACARE TWIN, Pg 33.
K. The AquaCare is now ready for use when required. Turn off the machine.
L. Always place back the handpiece in its holder after use.
The OPERATION section contains training procedures to get started, and
recommended settings for different uses.
If your observations do not match the results above, refer to the SERVICE section for
a full test procedure and troubleshooting tips. Alternatively, refer the problem to a
Velopex Representative or Authorised Distributor.
Use only original Velopex consumable products (powders, handpieces etc.). Other
makes may damage the AquaCare or negatively alter performance.