AquaCare Twin: Installation
Connecting the air supply
4 - 7 bar
See PARTS AND FEATURES: REAR VIEW OF AQUACARE TWIN, Pg 11.
The practice’s existing air supply system/compressor can be used provided it is fitted
with a water removing filter and shut-off valve in the supply line.
Medivance Instruments Ltd can supply an air filter (See ACCESSORIES, Pg 39).
Connect the blue air supply tube to the machine by pushing it fully into the blue
connector. See page 40 regarding installing an inline filter
Air Supply Requirements
Clean dry air to ISO 8573 Class 1.4.1.
Pressure between 4 bar (72psi) and 7 bar (100psi).
Maximum flow rate +70 l/min. 6mm O/D input connection.
Air
AIR
FOOT PEDAL