3.3.2 THE ELECTRICAL FUNCTIONS DO NOT 
WORK

• 

Does the relay click when you press the switch?

No: 

Check if the fuse has blown (does not apply 

to VELA Tango 300E/310E). The location of the 
fuse is shown on the spare parts drawing for the 
specific model. If the fuse has blown, replace the 
fuse and check all wires thoroughly, as well as all 
the electrical functions of the chair, e.g. up, down, 
and charge. If the fuse has not blown, check that the 
chair battery is charging properly (see 

section 3.3.1

). 

If the battery is charging properly, charge it for a 
longer time, e.g. overnight. 
If it does not charge properly, check that all wires 
are connected as shown in the spare parts drawing. 
If it still does not charge properly, contact VELA/
VELA Medical.

If the chair battery is fully charged, but the 
functions are still not working, 

replace the switch, 

spiral cable, motor control or the entire actuator. 
The motor control unit is an integral part of the 
actuator. Otherwise, contact VELA/VELA Medical.

Yes: 

Check that the chair battery is charging 

properly (see

 section 3.3.1

). If the battery is charging 

properly, charge it for a longer time, e.g. overnight. 
If it does not charge properly, contact VELA/VELA 
Medical. 

If the battery is fully charged, but the functions are 
still not working, replace the battery, spiral cable, 
motor control or the entire actuator, or contact 
VELA/VELA Medical.

3.3.3 THE ELECTRIC LIFT GOES UP BUT WON’T 
GO DOWN 

Activate the end switch for downward lift function 
by turning the adjusting screw a few times. 
The adjusting screw is a galvanised screw that 
requires a Philips screwdriver. It is located right next 
to the column, near the bottom side of the seat 
mechanism. See image:

(Does not apply to the VELA Tango 300E/310E)

3.3.4 THE ELECTRIC LIFT GOES DOWN BUT 
WON’T GO UP  

Check that the chair battery is charging properly 
(see 

section 3.1.1

). Check the voltage across the 

batteries. Voltage must be at least 24V on a fully 
charged battery. If it isn’t, replace the battery. 

If this does not solve the problem, replace the 
switch, spiral cable, motor control or the entire 
actuator, or contact VELA/VELA Medical.

3.4 REPLACING COMPONENTS

Regular service and maintenance may require 
parts of the chair to be disassembled to inspect 
individual components and provide access to other 
components, and to replace spare parts. Detailed 
assembly instructions with text and images are 
available on www.vela.eu, where they are divided 
into model-specific instructions, which include both 
standard configurations and accessories. Some 
instructions also include video tutorials.

3.4.1 RECOMMENDED TOOLS

Only a small number of tools are required to 
disassemble a VELA/VELA Medical chair. As a 
minimum, it is recommended to always have the 
following tools available in the toolbox: 

•  Allen key sizes 4, 5, and 6
•  Small, flat-bladed screwdriver
•  Philips screwdriver 
•  Shifting spanner up to 24 mm
•  Fibre hammer/rubber hammer
•  Lubricant, e.g. HSS2000

Several of the screws on VELA/VELA Medical chairs 
are fitted with locking glue from the factory to 
provide optimum safety for use. Therefore, only 
use original screws when replacing components to 
preserve locking glue in the intended locations.

3.5 ADJUSTING COMPONENTS  

REMEMBER:

 VELA/VELA Medical chairs are medical 

devices and the chair is customised for the individual 
user. Pay special attention to the position settings of 
various parts, such as the armrests, before carrying 
out any work on them. An occupational therapist 
may have decided on a precise setting for the user, 
so settings should always be set as they were before 
maintenance.

Содержание Tango

Страница 1: ...VELA Samba VELA Latin VELA Tango VELA Salsa VELA Tango VELA Salsa VELA Latin VELA Samba VELA VELA Medical S E R V I C E M A N U A L EN VERSION...

Страница 2: ...backrest 5 3 2 6 Armrest 5 3 2 7 Accessories 5 3 3 TROUBLESHOOTING CHAIRS WITH ELECTRIC LIFT 5 3 3 1 Recharging the chair battery 5 3 3 2 The electrical functions do not work 6 3 3 3 The electric lif...

Страница 3: ...s www vela eu www vela medical com VELA s customer service can also be contacted phone 45 96 34 76 00 or email mail vela eu mail vela medical com for further help 1 1 RIGHT OF COMPLAINT We grant the r...

Страница 4: ...over the chair The following sections contain further details about the checklist section 3 1 3 1 CHECKLIST FOR TESTING AND CHECKING 1 Frame 4 Brake function 7 Upholstery 2a Column gas spring 5 Seat m...

Страница 5: ...SEAT MECHANISM The seat mechanism is located under the seat and controls the seat tilt feature The release levers are also mounted on this plate Inspect the seat mechanism for excessive wear and tear...

Страница 6: ...e Does not apply to the VELA Tango 300E 310E 3 3 4 THE ELECTRIC LIFT GOES DOWN BUT WON T GO UP Check that the chair battery is charging properly see section 3 1 1 Check the voltage across the batterie...

Страница 7: ...ur website www vela eu 3 5 3 ADJUSTING ARMREST AND OTHER FUNCTIONS Several of the chair adjustment options are made with hand screws so the user can set the function without using tools If the user is...

Страница 8: ...kV Power Port Line to Earth and Neutral to Earth 0 5 kV 1 kV 2 kV IEC 61000 4 6 2013 Immunity to Conducted disturbances induces by RF fields Range 150kHz to 80MHz Outside ISM band 80 AM at 1kHz 3 VRM...

Отзывы: