2.0 CLEANING

VELA/VELA Medical chairs withstand thorough 
cleaning and can be reused. We recommend that 
professionals examine the chair following the 
checklist in 

section 3.1

.

 Several of the items on the 

list are described in further detail in 

section 3.0.

For instructions concerning regular cleaning, 
please refer to the chair’s user manual, which is 
available on our website: 

www.vela.eu/www.vela-

medical.com

2.1 CHAIRS WITH GAS SPRING HEIGHT 
ADJUSTMENT

Chairs with gas spring can be cleaned either with a 
hard-wrung cloth or through chemical disinfection 
with approved disinfectants. 

We also recommend regular maintenance after 
cleaning. This includes lubrication of inner and 
outer tubes described in 

section 3.2.2.

 

2.2 CHAIRS WITH ELECTRIC HEIGHT 
ADJUSTMENT

Chairs with electric height adjustment can be 
cleaned either with a hard-wrung cloth or with 
chemical disinfection using approved disinfectants.

3.0 MAINTENANCE

The purpose of a maintenance check is to identify 
defective and worn parts and similar issues. We 
recommend an annual overhaul of the chair as 
well as an overhaul when a new user takes over the 
chair. The following sections contain further details 
about the checklist (

section 3.1

). 

3.1 CHECKLIST FOR TESTING AND CHECKING

1. Frame

4. Brake function

7. Upholstery

2a. Column, gas spring

5. Seat mechanism

9. Electrical accessories

2b. Column, electrical

6. Armrests

10. Cleaning

3. Castors

8. Accessories

Inspect weldings 

Inspect for factures 

Tighten screws

Inspect brake function 

Inspect weldings

Listen for bad sounds 

Tighten screws

Inspect function of body support

Inspect function of push bar 

Inspect function of head support

Inspect function of leg supports 

Inspect function of thigh support

Inspect wires for damage 

Lubrication of moving parts 

Tighten screws 

Listen for bad sounds 

Charging function

Inspect rolling resistance 

Inspect bearings for backlash 

Tighten screws

Remove built-up dirt

Clean castors

Check function

Lubrication of moving parts  

Inspect for wear

Tighten screws

Inspect all functions 

Listen for bad sounds 

Tighten screws

Lubricate moving parts (not the 

lamellae)

Inspect function 

Inspect weldings

Inspect for fractures 

Tighten screws        

Inspect surface/upholstery

Inspect seat unholstery 

Inspect backrest upholstery

Inspect headrest upholstery 

Inspect body support upholstery 

Inspect loose seat covers

Electric brakes   

Switches  

Hand control unit

Upholstery 

Frame 

Armrests  

Accessories

Содержание Tango

Страница 1: ...VELA Samba VELA Latin VELA Tango VELA Salsa VELA Tango VELA Salsa VELA Latin VELA Samba VELA VELA Medical S E R V I C E M A N U A L EN VERSION...

Страница 2: ...backrest 5 3 2 6 Armrest 5 3 2 7 Accessories 5 3 3 TROUBLESHOOTING CHAIRS WITH ELECTRIC LIFT 5 3 3 1 Recharging the chair battery 5 3 3 2 The electrical functions do not work 6 3 3 3 The electric lif...

Страница 3: ...s www vela eu www vela medical com VELA s customer service can also be contacted phone 45 96 34 76 00 or email mail vela eu mail vela medical com for further help 1 1 RIGHT OF COMPLAINT We grant the r...

Страница 4: ...over the chair The following sections contain further details about the checklist section 3 1 3 1 CHECKLIST FOR TESTING AND CHECKING 1 Frame 4 Brake function 7 Upholstery 2a Column gas spring 5 Seat m...

Страница 5: ...SEAT MECHANISM The seat mechanism is located under the seat and controls the seat tilt feature The release levers are also mounted on this plate Inspect the seat mechanism for excessive wear and tear...

Страница 6: ...e Does not apply to the VELA Tango 300E 310E 3 3 4 THE ELECTRIC LIFT GOES DOWN BUT WON T GO UP Check that the chair battery is charging properly see section 3 1 1 Check the voltage across the batterie...

Страница 7: ...ur website www vela eu 3 5 3 ADJUSTING ARMREST AND OTHER FUNCTIONS Several of the chair adjustment options are made with hand screws so the user can set the function without using tools If the user is...

Страница 8: ...kV Power Port Line to Earth and Neutral to Earth 0 5 kV 1 kV 2 kV IEC 61000 4 6 2013 Immunity to Conducted disturbances induces by RF fields Range 150kHz to 80MHz Outside ISM band 80 AM at 1kHz 3 VRM...