6

VEGAVIB 61, 62, 63

53863-EN-170828

EPL Ga/Gb instrument

Temperature class Ambient tem-

perature on the 

electronics

Ambient tempera-

ture on the sensor
VEGAVIB 62

Permissible ambi-

ent temperature on 

the sensor with-

out temperature 

adapter
VEGAVIB 61/63

Permissible ambi-

ent temperature 

on the sensor 

with temperature 

adapter
VEGAVIB 61/63

T6

-40 … +55 °C

-20 … +54 °C

-50 … +85 °C

-50 … +85 °C

T5

-40 … +70 °C

-20 … +60 °C

-50 … +100 °C

-50 … +100 °C

T4

-40 … +80 °C

-20 … +60 °C

-50 … +135 °C

-50 … +135 °C

T3

-40 … +80 °C

-20 … +60 °C

-50 … +150 °C

-50 … +200 °C

T2, T1

-40 … +80 °C

-20 … +60 °C

-50 … +150 °C

-50 … +250 °C

Sect. 6.4.2 of IEC 1127-1 is taken into account with the stated permissible ambient temperatures on 

the sensor of VEGAVIB 62 and 66.
If the sensors of VEGAVIB VB6*.C****Z**** are operated at temperatures higher than those speci

-

fied in the above table, please make sure through appropriate measures that there is no danger of 

ignition from the hot surfaces. The max. permissible temperature on the electronics/housing should 

not exceed the values specified in the above table. The application conditions when operating in the 

absence of explosive mixtures can be found in the manufacturer information.
When the sensor of VEGAVIB 61, 63, 65 and 67 is operated in hazardous atmospheres of zone 0, 

there is no danger of ignition if it is operated under non-atmoshperic pressures from -1 to 16 bar and 

temperatures according to the temperature classes T6 … T1.

EPL Gb instrument

Temperature class Ambient tem-

perature on the 

electronics

Ambient tempera-

ture on the sensor
VEGAVIB 62

Permissible ambi-

ent temperature on 

the sensor with-

out temperature 

adapter
VEGAVIB 61/63

Permissible ambi-

ent temperature 

on the sensor 

with temperature 

adapter
VEGAVIB 61/63

T6

-40 … +55 °C

-20 … +70 °C

-50 … +85 °C

-50 … +85 °C

T5

-40 … +70 °C

-20 … +80 °C

-50 … +100 °C

-50 … +100 °C

T4

-40 … +80 °C

-20 … +80 °C

-50 … +135 °C

-50 … +135 °C

T3

-40 … +80 °C

-20 … +80 °C

-50 … +150 °C

-50 … +200 °C

T2, T1

-40 … +80 °C

-20 … +80 °C

-50 … +150 °C

-50 … +250 °C

If the VEGAVIB VB6*.C****Z**** are operated at temperatures higher than those specified in the 

above table, please make sure through appropriate measures that there is no danger of ignition 

from the hot surfaces. The maximum temperature on the electronics/housing should not exceed 

the values specified in the above table. The permissible operating temperatures and pressures are 

stated in the manufacturer information.

Permissible process pressure
EPL Ga instrument

For temperatures on the sensor according to the temperature classes T6 … T1, pressures under 

atmospheric conditions of 0.8 to 1.1 bar are permissible.

Содержание VEGAVIB 62

Страница 1: ...Safety instructions VEGAVIB 61 62 63 Intrinsic safety IECEx PTB 07 0015 X Two wire Document ID 53863...

Страница 2: ...ic electricity 7 6 Use of an overvoltage arrester 7 7 Impact and friction sparks 7 8 Potential equalisation 7 9 Installation 7 10 Material resistance 8 11 Locking mechanism of housing cover 8 12 Cable...

Страница 3: ...ally be observed The installation of explosion endangered systems must always be carried out by qualified person nel EPL Ga instrument The VEGAVIB VB6 C Z are installed in hazardous areas requiring a...

Страница 4: ...20 80 C E Cable FEP detection of solids in water 40 100 C K Cable PUR with Carbocer coating less buildup no corrosion abrasion protection 20 80 C L Cable FEP with Carbocer coating less buildup no cor...

Страница 5: ...Gb the intrinsically safe power supply and signal circuit must correspond to protection class ia For applications requiring instruments of type EPL Ga resp EPL Ga Gb the VEGAVIB VB6 C Z is preferably...

Страница 6: ...sor of VEGAVIB 61 63 65 and 67 is operated in hazardous atmospheres of zone 0 there is no danger of ignition if it is operated under non atmoshperic pressures from 1 to 16 bar and temperatures accordi...

Страница 7: ...ring operation maintenance and cleaning process related electrostatic charges e g by measuring media flowing past are ruled out 6 Use of an overvoltage arrester If necessary the VEGAVIB VB6 C Z can be...

Страница 8: ...stainless steel housing the lid has to be screwed into the stop and secured by screwing out one of the lid locking screws before setup of the instrument 12 Cable entries The supplied cable entry is s...

Страница 9: ...9 Notes VEGAVIB 61 62 63 53863 EN 170828...

Страница 10: ...10 Notes VEGAVIB 61 62 63 53863 EN 170828...

Страница 11: ...11 Notes VEGAVIB 61 62 63 53863 EN 170828...

Страница 12: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...

Отзывы: