6 Set up with the display and adjustment module
VEGAPULS SR 68 • 4 … 20 mA/HART two-wire
38294-EN-130624
Enter the requested parameters via the appropriate keys, save your
settings and jump to the next menu item with the
[ESC]
and
[->]
key.
Caution:
Note the following if the instrument with corresponding approval is
used as part of an overfill protection system according to WHG:
If a linearization curve is selected, the measuring signal is no longer
necessarily linear to the filling height. This must be considered by the
user especially when adjusting the switching point on the limit signal
transmitter.
With a reset, certain parameter adjustments carried out by the user
are reset.
The following reset functions are available:
Delivery status:
Restoring the parameter settings at the time of ship-
ment from the factory incl. the order-specific settings. A created false
signal suppression, user-programmable linearization curve as well as
the measured value memory will be deleted.
Basic settings:
Resetting the parameter settings incl. special pa-
rameters to the default values of the respective instrument. A created
false signal suppression, user programmable linearization curve as
well as the measured value memory will be deleted.
Setup:
Resetting of the parameter settings to the default values of the
respective instrument in the menu item Setup. User-generated false
signal suppression, user-programmed linearization curve, measured
value memory as well as event memory remain untouched. The
linearization is set to linear.
False signal suppression:
Deleting a previously created false signal
suppression. The false signal suppression created in the factory
remains active.
Peak values measured value:
Resetting of the measured min. and
max. distances to the actual measured value.
The following table shows the default values of the instrument. De-
pending on the instrument version, not all menu items are available or
differently assigned:
Additional adjustments
- Reset