50
13 Supplement
VEGAPULS C 22 • SDI-12
58345-EN-200806
Read temperature unit
Command
Response
Description
aXRTU!
a<value1><CR><LF>
a: sensor address
<value1>: temperature unit +0 = unit in [°C], +1 = unit in
[F], +2 = unit in [K]
<CR><LF>
Example:
Command
Response
Description
aXRTU!
0+0<CR><LF>
Write temperature unit
Command
Response
Description
aXWTU<value1>! a<value1><value2> <CR><LF>
a: sensor address
<value1>: temperature unit +0 = unit in [°C], +1 = unit in
[F], +2 = unit in [K]
<value2>: VVO-eee
<CR><LF>
Example:
Command
Response
Description
0XWTU+0!
0+0+000<CR><LF>
Valid data
0XWDU+4!
0+0+136<CR><LF>
No valid data
Current value is returned with a status 136
Read stage reference
Command
Response
Description
aXRSR!
a<value1><CR><LF>
a: sensor address
<value1>: stage refss.sss [m], +ss.sss [ft],
+ddddd [mm], sss.ss [in]
<CR><LF>
Example:
Command
Response
Description
aXRSR!
0+11.000<CR><LF>