30
11 Removal with glue connection
VEGAPULS Air 23 • Autarkic device with measured value transmission via radio technology
64314-EN-230606
11 Removal with glue connection
11.1 Dismounting instructions
The device can be removed from the surface of an IBC container by
unscrewing the sensor housing from the mounting ring. The mounting
ring with its adhesive connection remains on the surface of the vessel
and can be used for remounting the sensor.
If the mounting ring should be removed, follow the mounting steps in
the next section.
1
2
3
Fig. 20: Removal with glue connection
1 Vessel top
2 Mounting ring with glue surface
3 Sensor
Danger:
If the sensor shows obvious damage, there is a risk of ignition due to
a possibly damaged Lithium cell. In this case, the device must not be
packed or transported any further.
11.2 Dismounting steps mounting ring
The following tools are required to remove the mounting ring:
•
Strap wrench
•
if necessary a wooden scraper
Danger:
Depending on the medium in the IBC container, damage to the con-
tainer may pose further risks for the user, the effects of which cannot
be foreseen. Therefore, inform yourself about the contents of the IBC
container before starting dismantling work and follow the instructions
of the safety data sheet for the medium.
Caution:
The use of pointed or levering tools for dismantling may result in dam-
age to the IBC container or equipment with the consequences shown
above. Therefore, follow the procedure described below and use only
tools recommended by VEGA.
Proceed as follows while dismounting:
Tool
Dismount