9
VEGAPULS 6X
66216-EN-220518
•
If grounding of the cable screening is necessary, this must be carried out acc. to the valid stand-
ards and regulations, e.g. acc. to IEC/EN 60079-14.
•
Close the housing lid (s) up to the stop before starting operating, to ensure the IP protection rat
-
ing specified on the type label.
Cable glands, threaded openings
Type
Thread
Cable diameter [mm]
Torques [Nm]
Hummel EXIOS A2F 1.608.2003.50
M20 x 1.5
6 … 12 mm
8
Hummel EXIOS A2F 1.608.1203.70
½ NPT
6 … 12 mm
8
Hummel EXIOS MZ 1.6Z5.2000.51
M20 x 1.5
9 … 13 mm
8
Hummel EXIOS MZ 1.6Z5.1200.70
½ NPT
9 … 13 mm
8
The specified torques are test torques and can only be regarded as reference values. The manufac
-
turer's mounting instructions provided must be observed.
If suitable cable glands or cable insertion possibilities not included in the scope of supply are used,
these must be compatible with the threaded openings.
Aluminium housing with M20 x 1.5 thread, ½ NPT thread
Stainless-steel housing (fine cast) with M20 x 1.5 thread, ½ NPT thread
Maintenance
To ensure the functionality of the device, periodic visual inspection is recommended for:
•
Secure mounting
•
No mechanical damages or corrosion
•
Worn or otherwise damaged cables