16

5 Set up with the display and adjustment module

VEGAPULS 69 • 4 … 20 mA/HART - four-wire

47255-EN-140801

Information:

This is generally recommended, however, for use of the Asset 

Management functions it is absolutely necessary. Saving should be 

carried out with a very low level.
The function "

Echo curve memory

" allows up to ten individual echo 

curves to be stored, for example to detect the measurement behav-

iour of the sensor in different operating conditions.

With the adjustment software PACTware and the PC, the stored echo 

curves can be displayed with high resolution and used to recognize 

signal changes over time. In addition, the echo curve saved during 

setup can also be displayed in the echo curve window and compared 

with the actual echo curve.

The following circumstances cause interfering reflections and can 

influence the measurement:

• 

High sockets

• 

Vessel installations such as struts

• 

Agitators

• 

Buildup or welded joints on vessel walls

Note:

A false signal suppression detects, marks and saves these false 

signals so that they are no longer taken into account in the level 

measurement.
This should be done with a low level so that all potential interfering 

reflections can be detected.

Proceed as follows:
1.  Select with 

[->]

 the menu item "

False signal suppression

" and 

confirm with 

[OK]

.

2.  Confirm again with 

[OK]

.

3.  Confirm again with 

[OK]

.

Additional adjustments - 

False signal suppression

Содержание VEGAPULS 69

Страница 1: ...Quick setup guide Radar sensor for continuous level measurement of bulk solids VEGAPULS 69 4 20 mA HART four wire Document ID 47255...

Страница 2: ...r housing 11 5 Set up with the display and adjustment module 5 1 Insert display and adjustment module 13 5 2 Parameter adjustment Quick setup 13 5 3 Parameter adjustment Extended adjustment 15 6 Suppl...

Страница 3: ...t instrument complying with all prevailing regulations and guidelines The instrument must only be operated in a technically flawless and reliable condition The operator is responsible for the trouble...

Страница 4: ...ets the LPR Level Probing Radar radio standard EN 302729 1 2 It is approved for unrestricted use inside and outside of closed ves sels in countries of the EU and EFTA that have implemented this standa...

Страница 5: ...safety VEGAPULS 69 4 20 mA HART four wire 47255 EN 140801 Please help us fulfill this obligation by observing the environmental instructions in this manual Chapter Packaging transport and storage Chap...

Страница 6: ...12 Data Matrix Code for Smartphone App 13 Symbol of the device protection class 14 ID numbers instrument documentation 15 Reminder to observe the instrument documentation 16 Notified authority for CE...

Страница 7: ...4 20 mA HART four wire 47255 EN 140801 Download the smartphone app VEGA Tools from the Apple App Store or the Google Play Store Scan the Data Matrix code on the type label of the instrument or Enter t...

Страница 8: ...0 90 and 180 Double chamber housing Angle of inclination 90 infinitely variable Angle of inclination in two steps 0 and 90 3 2 Mounting instructions Radar sensors for level measurement emit electroma...

Страница 9: ...radar sensor on the vessel top If you cannot maintain this distance you should carry out a false signal storage during setup This applies particularly if buildup on the vessel wall is expected In suc...

Страница 10: ...ew the housing cover 2 Loosen compression nuts of the cable glands 3 Remove approx 10 cm 4 in of the cable mantle signal output strip approx 1 cm 0 4 in insulation from the ends of the indi vidual wir...

Страница 11: ...n The electrical connection is finished Information The terminal blocks are pluggable and can be removed from the housing insert To do this lift the terminal block with a small screwdriv er and pull i...

Страница 12: ...on IS GND 5 1 L N 2 6 7 8 Fig 6 Connection compartment with double chamber housing with low voltage Terminal Function Polarity 1 Voltage supply L 2 Voltage supply N 5 4 20 mA output active 6 4 20 mA o...

Страница 13: ...in the desired position and turn it to the right until it snaps in 3 Screw housing cover with inspection window tightly back on Disassembly is carried out in reverse order The display and adjustment...

Страница 14: ...this menu item you select the type of bulk solid The selection com prises different granulate sizes 3 Application Vessel bottom In this menu item you specify the application and the form of the ves s...

Страница 15: ...adjustment 1 Min level max meas distance 2 Max level min meas distance 3 Reference plane If these values are not known an adjustment with the distances of for example 10 and 90 is possible Starting po...

Страница 16: ...signal changes over time In addition the echo curve saved during setup can also be displayed in the echo curve window and compared with the actual echo curve The following circumstances cause interfe...

Страница 17: ...signal suppression has already been created in the sensor the following menu window appears when selecting False signal suppression Delete An already created false signal suppression will be com plete...

Страница 18: ...1 5 or NPT Wire cross section spring loaded terminals Massive wire stranded wire 0 2 2 5 mm AWG 24 14 Stranded wire with end sleeve 0 2 1 5 mm AWG 24 16 Voltage supply Operating voltage Version for lo...

Страница 19: ...19 Notes VEGAPULS 69 4 20 mA HART four wire 47255 EN 140801...

Страница 20: ...of delivery application practical use and operat ing conditions of the sensors and processing systems correspond to the information available at the time of printing Subject to change without prior n...