4
E
nter the suitable distance value in m for the empty vessel
(
e
.
g
.
distance from the sensor to the vessel bottom
)
corresponding to
the percentage value
.
5
S
ave settings with
[
OK
]
and move with
[
ESC
]
and
[-
>
]
to the max
.
adjustment
.
P
roceed as follows
:
1
S
elect with
[-
>
]
the menu item max
.
adjustment and con
fi
rm with
[
OK
]
.
2
P
repare the percentage value for editing with
[
OK
]
and set the
cursor to the requested position with
[-
>
]
.
3
S
et the requested percentage value with
[
+
]
and save with
[
OK
]
.
T
he cursor jumps now to the distance value
.
4
E
nter the appropriate distance value in m
(
corresponding to the
percentage value
)
for the full vessel
.
K
eep in mind that the max
.
level must lie below the min
.
distance to the antenna edge
.
5
S
ave settings with
[
OK
]
M
in
.
and max
.
measured values are saved in the sensor
.
T
he values
are displayed in the menu item
"
P
eak values
"
.
W
hen non
-
contact level sensors are used
,
the measurement can be
in
fl
uenced by the respective process conditions
.
I
n this menu item
,
the
measurement reliability of the level echo is displayed as d
B
value
.
T
he
measurement reliability equals signal strength minus noise
.
T
he higher
the value
,
the more reliable the measurement
.
W
ith a functioning
measurement
,
the values are
>
10
d
B
.
S
etup
/
M
ax
.
adjustment
D
iagnosis
/
P
eak value
D
iagnosis
/
M
easurement
reliability
VEGAPULS
62
standpipe version
•
F
oundation
F
ieldbus
25
6
S
et up with the indicating and adjustment module
36510
-
EN
-
120301
Содержание VEGAPULS 62 standpipeversion
Страница 1: ...Operating Instructions VEGAPULS 62 standpipe version Foundation Fieldbus Document ID 36510 Radar...
Страница 73: ...VEGAPULS 62 standpipe version Foundation Fieldbus 73 Index 36510 EN 120301...
Страница 74: ...74 VEGAPULS 62 standpipe version Foundation Fieldbus Index 36510 EN 120301...
Страница 75: ...VEGAPULS 62 standpipe version Foundation Fieldbus 75 Index 36510 EN 120301...