9 Adjustment menu
VEGAPULS 11 • Two-wire 4 … 20 mA
58350-EN-221027
9 Adjustment menu
9.1 Menu overview
Device information
Actual measured values
Device status
Device name, software ver-
sion, serial number
Percent, filling height, distance, measurement reliability,
electronics temperature, meas. rate etc.
OK, error indication
Menu item
Selection
Basic settings
Measurement loop name
Alphanumeric characters
Sensor
Medium
Liquid
Bulk solid
Liquid
Application liquid
Storage tank, agitator tank, dosing tank, pumping sta-
tion/pump shaft, rain overflow basin, tank/collection
basin, plastic tank (measurement through tank top), mo-
bile plastic tank (IBC), level measurement in waters, flow
measurement flume/overflow, demonstration
Storage tank
Application bulk solid
Silo (slim and high), bunker (large volume), stockpile
(point measurement/profile detection), crusher, demon
-
stration
Silo (slender and
high)
Units
Distance unit of the device
Temperature unit of the instrument
Distance in m
Temperature in °C
Adjustment
Max. adjustment (distance A)
Min. adjustment (distance B)
Max. adjustment
0,000 m
Min. adjustment
8,000 m
Menu item
Selection
Basic settings
Damping
Integration time
0 s
Current output
Output characteristics
0 … 100 % corre-
spond to 4 … 20 mA
Current range
3.8 … 20.5 mA
Reaction when malfunctions occur
< 3.6 mA
Linearisation
Linearization type
Linear
Intermediate height
Scaling
Scaling size
Scaling unit
Scaling format
100 % correspond to
0 % correspond to
Volume
l
100 l
0 l
Start image
Basic functions
Extended functions