7
I
nsert the wire ends into the open terminals according to the wiring
plan
F
ig
.
10
:
C
onnection steps
6
and
7
8
P
ress down the opening levers of the terminals
,
you will hear the
terminal spring closing
9
C
heck the hold of the wires in the terminals by lightly pulling on
them
10
C
onnect the screen to the internal ground terminal
,
connect the
outer ground terminal to potential equalisation
11
T
ighten the compression nut of the cable entry
.
T
he seal ring must
completely encircle the cable
12
S
crew the housing cover on
T
he electrical connection is
fi
nished
.
5
.
3
W
iring plan
,
single chamber housing
T
he following illustrations apply to the non
-
E
x as well as to the
E
x
-
ia
version
.
VEGAFLEX
61
•
P
ro
fi
bus
PA
19
5
C
onnecting to power supply
31835
-
EN
-
100426