background image

VEGACOM 557 - without protocol

9

Technical data

4 Technical data

Power supply

Operating voltage

U

nom

= 24 V AC (20 … 53 V), 50/60 Hz or
= 24 V DC (20 … 72 V)

Power consumption

approx. 6 VA

Fuse

1 A, slow-blow

Electrical connection

Component

multiple plug acc. to DIN 41 612, series F
48-pole (d, b, z) with coding holes

Module in carrier
BGT 596 or BGT LOG 571

suitable multipoint connector acc. to DIN 41 612
with connection via standard technologies

Housing type 505

via screw terminals max. 1 x 1,5 mm

2

Indicating elements

LED in front plate

green BA: without function
red (flashing): DISBUS/LOGBUS-failure
red (flashing irregularly): no PC/PLC-outputs
co-ordinated to the measurement loop
red (permanently): failure
green on: operating voltage on

Meas. data input DISBUS

Data transmission

DISBUS (digital data transmission)

Connection line

2-wire standard line (screening recommended)

Line length

max. 1000 m

Meas. data input LOGBUS

Data transmission

LOGBUS (digital data transmission)

Connection line

connection via BUS-plug

PC-interface

Interface standard

RS 232 C

Line length

max. 15 m

Transmission rate in baud 

1)

300, 600, 1200, 2400, 4800, 9600, 19200, 38400

Transmission format 

1)

8 data bits, 1 Stop bit, no parity or even parity

Plug in the front plate

SUB-D-plug  connector,  9-pole,  pins

1)

 via DIL-switch adjustable on the basic board

Содержание VEGACOM 557

Страница 1: ...in in out out Level and Pressure Operating Instructions VEGACOM 557 PC MODEM Interface on 557 BA PC ...

Страница 2: ...ote Ex area Please note the approval documents attached yellow binder and especially the included safety data sheet Contents Safety information 2 Note Ex area 2 1 Product description 1 1 Configuration 4 1 2 Functional principle 5 1 3 Function 6 2 Adjustment 2 1 Configuration and parameter adjustment of connected measuring systems 7 2 2 Visualisation 7 2 3 Remote parameter adjustment 7 3 Measuring ...

Страница 3: ...sing 13 6 Electrical connection 6 1 Connection instructions 14 6 2 Connection plans 14 7 1 Switch positions on VEGACOM 557 15 7 Parameter adjustment of the interface 15 8 VVO or VV in conjunction with VEGACOM 557 16 9 Set up 9 1 Set up checklist 16 10 Connection examples 10 1 PC on the front interface 17 10 2 Modem on front RS232 interface 17 ...

Страница 4: ...made directly via the cable on the RS232 interface of VEGACOM 557 It is possible to provide a connection via modem 1 1 Configuration VEGACOM 557 is designed as a component in 19 technology with 5 TE width 1 TE 5 08 mm acc to DIN 41 494 It can be used in carrier BGT 596 in VEGALOG carrier BGT LOG 571 in housing type 505 The electrical connection is made via a plug connector acc to DIN 41 612 on the...

Страница 5: ...n 0 10 5 2 0 10 5 521 Ex on 2 on on on PC on on VEGALOG 1 2 3 4 5 6 7 8 9 10 on 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 on 1 2 3 4 5 6 7 8 9 10 on 571 CPU VEGALOG 571 EA VEGALOG 571 EX VEGALOG 571 AT VEGALOG 571 AR VEGALOG 571 AT VEGALOG 571 AR VEGALOG 571 AD on 557 BA PC on 557 BA PC VEGAMET VEGAMET VEGASEL VEGASEL VEGATOR VEGATOR VEGATOR VEGACOM VEGACOM VEGASTAB 593 on CONNECT VEGASTAB 593 on ...

Страница 6: ...f the Visual VEGA VV visu alisation software By means of the VEGA Visual Operating VVO parameter adjustment or configuration of connected level and pressure measuring systems is possible The communication can be made via a PC or modem to the remote PC remote parameter adjustment directly connected to VEGACOM 557 The data communication between VEGACOM 557 and the PC is always initiated by the PC wi...

Страница 7: ...tment of connected measur ing systems An additional function of VEGACOM 557 is the adjustment of the connected measuring system The adjustment is made via a PC by means of the VEGA Visual Operating VVO adjustment software The PC is connected via an interlink cable to the 9 pole SUB D plug in the front plate of VEGACOM 557 The VEGA adjustment concept includes the convenient configuration and parame...

Страница 8: ...measurement loops to groups saving of fault signals and measured values re corder function Suitable for networks 3 Measuring system with digital communication 1 2 1 4 4 3 5 5 DISBUS LOGBUS Automatisation system 3 VEGACOM 557 Interface converter for conversion of VEGA specific protocols in standard for mats Suitable for connection to the DIS BUS output of VEGAMET series 500 600 signal conditioning ...

Страница 9: ...on red flashing DISBUS LOGBUS failure red flashing irregularly no PC PLC outputs co ordinated to the measurement loop red permanently failure green on operating voltage on Meas data input DISBUS Data transmission DISBUS digital data transmission Connection line 2 wire standard line screening recommended Line length max 1000 m Meas data input LOGBUS Data transmission LOGBUS digital data transmissio...

Страница 10: ... connection and appropriate interface reference voltage 250 V test voltage 2 kV CE approval conformity judgement VEGACOM 557 meets the protective regulations of EMVG 89 336 EWG and NSR 73 23 EWG The conformity has been judged acc to the following standards EMVG Emission EN 50 081 1 1993 Susceptibility EN 50 082 2 1995 NSR EN 61 010 1 1993 Ambient conditions Permissible ambient temperature 20 C 60 ...

Страница 11: ...VEGACOM 557 without protocol 11 162 5 5 15 100 on 557 BA 128 4 25 4 5 TE Technical data Multipoint connector 100 x 160 x 1 5 European size Multiple plug LOGBUS plug front RS232C interface ...

Страница 12: ... in the carrier BGT 596 or into the housing type 505 In conjunction with the LOGBUS the VEGA COM 557 is mounted in the carrier BGT LOG 571 The position is individually selectable the system adjusts automatically auto con figuration Coding A mechanical coding system avoids inter changing among the different module cards in the carrier or in the housing The coding system consists of three coding pin...

Страница 13: ...he multipoint connector is available as fol lows Wire Wrap standard connection 1 0 mm x 1 0 mm plug connection 2 8 mm x 0 8 mm Termi Point standard connection 1 6 mm x 0 8 mm soldering connection screw terminals 0 5 mm2 For mounting the module please note the operating instructions of the carrier Housing type 505 This housing is equipped with a multipoint connector Before mounting please check whe...

Страница 14: ...nection plans 1 2 3 6 7 8 9 5 4 RS 232 CTS RxD TxD RTS GND Pin Description I O 1 CTS clear to send I 2 RxD receive data I 3 TxD transmit data O 4 RTS request to send O 5 GND ground D SUB plug front plate Multipoint connector rear of instru ment Supply voltage DISBUS not used with VEGALOG Interface cable PC VEGACOM With an RS 232 interface cable interlink cable you connect the PC with the software ...

Страница 15: ...US Automatic Modem initialisation In position ON the modem will be automati cally initialised when connecting the modem to the VEGACOM interface Transmission rate For VVO or VV the transmission rate must be adjusted to 9600 baud 7 1 Switch positions on VEGACOM 557 A 6 pole DIL switch block for adjustment of the RS 232 PC interface in the front plate is located on the basic board Before inserting V...

Страница 16: ...the connection via the telephone net there is no difference between the adjustment e g via VVO and the adjustment when directly connected via cable In the further chapters you find the connection examples either for direct connection of the PC or for the connection via modem 9 Set up 9 1 Set up checklist 1 Hardware requirements VEGACOM 557 version PC Modem Interface PC directly via interlink cable...

Страница 17: ...n the basic board On LOGBUS this switch is meaningless 10 2 Modem on front RS232 interface In this example a PC is connected via MO DEM and the front interface of VEGACOM 557 to the VEGA processing system Two analogue modems are applied in order to use the telephone net for remote param eter adjustment With ISDN connection an additional so called terminal adapter a b converter is required taking o...

Страница 18: ...18 VEGACOM 557 without protocol Notes ...

Страница 19: ...VEGACOM 557 without protocol 19 Notes ...

Страница 20: ...sensors and processing systems correspond to the actual knowl edge at the date of printing Technical data subject to alteration 23290 EN 021119 VEGA Grieshaber KG Am Hohenstein 113 77761 Schiltach Germany Phone 49 0 7836 50 0 Fax 49 0 7836 50 201 E Mail info de vega com www vega com ...

Отзывы: