24
7 Diagnostics and servicing
VEGACAP 64 • Two-wire
30015-EN-230510
13. Check the hold of the wires in the terminals by lightly pulling on
them
14. Check cable gland on tightness. The seal ring must completely
encircle the cable.
15. Mount the probe into the vessel. Make sure that the probe is
uncovered.
2
1
Fig. 17: Compensation key
1 Measuring range selection switch (compensation key)
2 Control lamp
16. Keep the measuring range selection switch (1) pushed until the
control lamp (2) flashes green.
17. Carry out the adjustment again. See chapter "
Set-up, adjustment
elements
".
18. Screw the housing lid back on
The electronics exchange is now finished.
7.4 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
Proceed as follows in case of repair:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.