35
7 Set up with the integrated display and adjustment unit
VEGABAR 38 • Three-wire with IO-Link (2 x transistor or 4 … 20 mA plus 1 x transistor)
57532-EN-191014
Menu item code:
•
RES
Parameter:
•
Basic settings
•
Delivery status
7)
Basic settings:
Resets the parameter settings to the default values
of the respective device. The order-related settings are not transferred
to the current parameters after this reset.
Delivery status:
Resets the parameter settings to the delivery status.
Information:
The current status of the access protection, the Bluetooth access
code and the device code are not reset.
7.4.3 Diagnostics
In this menu item, the device status is displayed.
Menu item code:
•
STA
In the event of an error, the error code, e.g. F017, and an error de-
scription, e.g. "
Adjustment span too small
" are displayed.
This menu item displays the number of parameter changes made.
Menu item code:
•
PCO
Status
Parameter modification
counter
7)
Parameter delivery status only available with a parameter adjustment that
deviates from the basic settings, e.g. customer-specific adjustment