6 Set up with the display and adjustment module PLICSCOM
VEGAPULS 61 • Profibus PA
28444-EN-141215
The menu item "
Extended setting
" offers the possibility to optimise
VEGAPULS 61 for applications in which the level changes very
quickly. To do this, select the function "
Quick level change > 1 m/min.
".
Extended setting
quick level change > 1 m/min.
Note:
Since with the function "
Quick level change > 1 m/min.
" the genera-
tion of an average value of the signal processing is considerably
reduced, false reflections by agitators or vessel installations can
cause measured value fluctuations. A false echo memory is thus
recommended.
This function enables reading out parameter adjustment data as well
as writing parameter adjustment data into the sensor via the display
and adjustment module. A description of the function is available in
the operating instructions manual "
Display and adjustment module
".
The following data are read out or written with this function:
•
Measured value presentation
•
Adjustment
•
Medium
•
Standpipe inner diameter
2)
•
Vessel form
•
Damping
•
Linearisation curve
•
Sensor-TAG
•
Displayed value
•
Scaling unit (Out-Scale unit)
•
Positions after the decimal point (scaled)
•
Scaling PA/Out-Scale 4 values
•
Unit of measurement
•
Language
The following safety-relevant data are
not
read out or written:
•
Sensor address
•
PIN
Copy sensor data
Copy sensor data?
Basic adjustment
If the "
Reset
" is carried out, the sensor resets the values of the follow-
ing menu items to the reset values (see chart):
3)
Function
Reset value
Sensor address
126
Extended setting/Quick
level change
Copy sensor data
Reset
2)
With standpipe versions.
3)
Sensor-specific basic adjustment.