6 Set up with the display and adjustment module PLICSCOM
VEGAPULS 61 • Profibus PA
28444-EN-141215
If you want to measure the total height of both liquids reliably, please
contact our service department or use an instrument specially de-
signed for interface measurement.
In the main menu item "
Basic adjustment
", the individual submenu
items should be selected one after the other and provided with the
correct parameter values.
Start your parameter adjustment with the following menu items of the
basic adjustment:
Proceed as follows:
1. Move from the measured value display to the main menu by
pushing
[OK]
.
▶
Basic adjustment
Display
Diagnostics
Service
Info
2. Select the menu item "
Basic adjustment
" with
[->]
and confirm
with
[OK]
. Now the menu item "
Min. adjustment
" is displayed.
Min. adjustment
0.00 %
=
5.000 m(d)
4.000 m(d)
3. Prepare the % value for editing with
[OK]
and set the cursor to the
requested position with
[->]
. Set the requested percentage value
with
[+]
and save with
[OK]
. The cursor jumps now to the distance
value.
4. Enter the suitable distance value in m for the empty vessel (e.g.
distance from the sensor to the vessel bottom) corresponding to
the percentage value.
5. Save the settings with
[OK]
and move to "Max. adjustment" with
[->]
.
Proceed as follows:
Max. adjustment
100.00 %
=
1.000 m(d)
2.000 m(d)
1. Prepare the % value for editing with
[OK]
and set the cursor to the
requested position with
[->]
. Set the requested percentage value
with
[+]
and save with
[OK]
. The cursor jumps now to the distance
value.
2. Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. Keep in mind that the max.
level must lie below the dead band.
3. Save the settings with
[OK]
and move to "Medium selection" with
[->]
.
Carry out min. adjustment
Carry out max. adjust-