FX80 / FX81 PON Optical Power Meters User Manual
Page
16
of
50
The FX80/FX81 is now ready to measure PON levels and compare values against the Pass/Fail
thresholds settings. The thresholds settings will remain active until changed.
If the FX80/FX81 battery is allowed to fully discharge, all settings such as
Pass/Fail and Device settings will return to original factory defaults. To
retain Pass/Fail settings for future use, save them to a profile.
6.2.2.2 Creating a Pass/Fail Threshold Profile for Measurements
One profile with threshold values can be created and saved in memory for future recall.
To save the current set of thresholds as a profile:
1. In the P/F mode, enter the desired values, but do NOT exit the
Setting Pass/Fail
thresholds
mode.
2. To save the values as the profile, press
|F3| (Save)
.
3. To load the previously saved profile, press
|F2| (Load)
. The values from the custom
profile will appear immediately on the screen.
4.
To load the default threshold profile, press
|Down|
until
Load Default
is selected,
then press
|Shift/Enter|
. The values from the default profile appear immediately on
the screen.
Selecting the Default profile, FX80