Entering Code 50 will display the version (x.x) of the 

meter. The display then returns to 

CodE 00

. Press the 

SEL

 

button to exit the module. 

Entering Code 66 will overwrite all user settings with 

the factory settings. The meter will display 

rESEt

 and then 

return to 

CodE 00

. Press the 

SEL

 button to exit the module. 

8

6.2  Module 2 - S

econdary

 f

unction

 p

araMeterS

 (

2-SEC

)

SEL

Max Capture 

Delay Time

Max Display 

Enable

Min Display 

Enable

Access Code 

For Service 

Operations

Min Capture 

Delay TIme

Factory 

Service 

Operations

2-SEC

HI-En

HI-t

LO-En

LO-t

FCS

CodE

Pro

PARAMETER MENU

MIN DISPLAY ENABLE

USER INPUT ASSIGNMENT

Select the value(s) to which the User Input Function is assigned. The User 

Input Assignment only applies if a selection of reset, display hold, or print and 

reset is selected in the User Input Function menu.

 MODE

DISPLAY

No Function

NO

DESCRIPTION

User Input disabled.

Program Mode Lock-out

P-Loc

Zero Input  

(Edge triggered)

ZErO

Zero the Input Display value causing 

Display Reading to be Offset.

COLOr

d-SEL

d-LEV

d-HLd

rESEt

Resets the assigned value(s) to the 

current input value.

HI-LO 

HI

dSP

LO

See Programming Mode Access chart 

(Module 3).

YES

NO

 MODE

DESCRIPTION

DISPLAY

Setpoint 1 and 2 Reset

Setpoint 1 Reset
Setpoint 2 Reset

Print and Reset

Reset both setpoint 1 and 2 outputs.

Resets setpoint 1 output.
Resets setpoint 2 output.

Same as Print Request followed by a 

momentary reset of the assigned value(s).

rSt12

rSt-1

rSt-2

P-r5t

Print Request

Serial transmit of the active parameters 

selected in the Print Options menu 

(Module 5).

Print

2.0

HI-t

NO

LO-En

NO

FCS

MAX CAPTURE DELAY TIME

When the Input Display is above the present MAX value for the entered 

delay time, the meter will capture that display value as the new MAX reading. 

A delay time helps to avoid false captures of sudden short spikes.

2.0

LO-t

MIN CAPTURE DELAY TIME

When the Input Display is below the present MIN value for the entered delay  

time, the meter will capture that display value as the new MIN reading. A delay 

time helps to avoid false captures of sudden short spikes.

0.0

  

to 

 

999.9

  

seconds

USER INPUT FUNCTION

NO

USrIN

dSP

U-ASN

NO

HI-En

MAX DISPLAY ENABLE

0.0

  

to 

 

999.9

  

seconds

Select 

yES

 to perform either of the Factory Service Operations shown below.

FACTORY SERVICE OPERATIONS

yES

NO

YES

NO

Enables the Maximum Display Capture capability.

Enables the Minimum Display Capture capability.

The RFMP uses stored calibration values to provide 

accurate voltage and current measurements. Over time, 

the electrical characteristics of the components inside the 

meter could slowly change, therefore the stored calibration 

values may no longer accurately define the input circuit. 

For most applications, recalibration every 1 to 2 years should be sufficient.

Calibration of the RFMP involves a voltage or current calibration, which 

should only be performed by individuals experienced in calibrating electronic 

equipment. Allow a 30 minute warm up for eqiupment and unit before 

performing any calibration related procedures. The following procedures should 

be performed at an ambient temperature of 15 to 35°C (59 to 95°F). 

CAUTION: The accuracy of the calibration equipment will directly affect the 

accuracy of the RFMP.

Calibration

1.  Connect the negative lead of a precision DC source with an accuracy of 0.01% 

or better to the COMM. Leave the positive lead of the DC source unconnected.

2.  With the display at 

CodE 48

, press and hold the 

SEL

 button for 2 seconds. Unit 

will display 

CAL NO

.

3.  Press the 

RST

 button to select the range to be calibrated.

4.  Press the 

SEL

 button. Display reads 

0.0A

 (

0.0 V

 for voltage).

5.  Apply 0 signal: 

Current: leave the positive lead of the DC source unconnected.

Voltage: apply a short to the input or connect the positive lead of the DC 

source to INP+ and set the source to 0.

 Press 

SEL

. Display reads 

CALC

 for about 8 seconds.

6.  When the display reads the selected range (10 V, 20 mA, or 50 mA), connect 

the positive lead of the DC source to INP+ and apply the full scale input signal 

for the range. Press 

SEL

. Display reads 

CALC

 for about 8 seconds.

7.  Repeat steps 3 through 6 for each input range to be calibrated. When display 

reads 

CAL NO

, press the 

SEL

 button to exit calibration.

CALIBRATION

48

CodE

66

CodE

50

CodE

RESTORE FACTORY DEFAULT SETTINGS

VIEW VERSION DISPLAY

Backlight Color  

(Edge Triggered)

Display Select  

(Edge Triggered)
Display Intensity Level  

(Edge Triggered)

Display Hold

Reset (Edge triggered)

Change backlight color with each 

activation (backlight version only).

Advance once for each activation.

Increase intensity one level for each 

activation (backlight version only).

Holds the assigned display, but all other 

meter functions continue as long as 

activated (maintained action).

Содержание RFMP Series

Страница 1: ...FLEX MINI can be powered from an optional Veeder Root Micro Line Sensor Power Supply MLPS which attaches directly to the back of a FLEX MINI The MLPS is powered from 85 to 250 VAC and provides up to 4...

Страница 2: ...C Green Display 1 2 35 to 75 C 3 35 to 65 C 4 35 to 50 C 5 35 to 35 C Storage Temperature 35 to 85 C Operating and Storage Humidity 0 to 85 max relative humidity non condensing Vibration to IEC 68 2 6...

Страница 3: ...5 SERIAL COMMUNICATIONS CARD Type RS485 multi point balanced interface non isolated Baud Rate 300 to 38 4k Data Format 7 8 bits odd even or no parity Bus Address 0 to 99 max 32 meters per line Transmi...

Страница 4: ...thin an enclosure should be routed as far as possible from contactors control relays transformers and other noisy components 5 In extremely high EMI environments the use of external EMI suppression de...

Страница 5: ...UNREG 400mA 12V DC COMMON M3424B TERMINAL AC INPUT MLPS1 2 WIRE TRANSMITTER COMM INP _ MODEL INPUT POWER OUTPUT POWER 85 250 VAC 50 60Hz 14VA OPERATING TEMP 0 TO 60 C 12VDC UNREG 400mA 12V DC COMMON...

Страница 6: ...the SEL button with Pro NO displayed This will commit any stored parameter changes to memory and return the meter to the Display Mode If power loss occurs before returning to the Display Mode verify...

Страница 7: ...Values have to be derived from the actual input signal source or simulator the Apply APLY scaling style must be used KEy APLY 0 000 INP 1 INPUT VALUE FOR SCALING POINT 1 For Key in KEY style enter th...

Страница 8: ...ow FACTORY SERVICE OPERATIONS yES NO YES NO Enables the Maximum Display Capture capability Enables the Minimum Display Capture capability The RFMP uses stored calibration values to provide accurate vo...

Страница 9: ...acklight Unit Only 3 dSP dSP t SEL rSt ZErO UNItS COLOr CodE Display Intensity Level d LEV ScroL Display Scroll Enable PARAMETER MENU The yES selection allows the SEL button to toggle through the enab...

Страница 10: ...ced are affected by the hysteresis When the setpoint is a control output usually balanced hysteresis is used For alarm applications usually unbalanced hysteresis is used For unbalanced hysteresis mode...

Страница 11: ...s Refer to the RFMCOMUSB bulletin for details on the FLEX MINI USB programming and programming requirements PARAMETER MENU button or user input manual reset serial reset command or meter power cycle W...

Страница 12: ...tOF n rSt n rEn n ChC n Standby Operation Stb n Front Panel Reset Enable Front Panel Display Select Enable Display Update Time dSP t SEL rSt 3 dSP Display Intensity Level COLOr Units Indicator Select...

Отзывы: