Troubleshooting Guide

HydrX Quick Help - ATG Control

Section 5 - HydrX System Alarm Troubleshooting

g. On the IN OUT screen (see Figure 13) in the D_OUT number buttons, touch:

- 0-to energize the inlet valve
- 1-to energize the outlet valve
- 2-to energize the pump hook 
- 3-To test the FCC alarm

Figure 13.   IN OUT Screen

Pressure Fault Actions

The HydrX system requires a minimum15 psi reading on its pressure sensor for system startup with new filters. If pressure is 

lower than 15 psi the following steps can be taken.
1. Enter the FCC Self Diag. screen (see Figure 11) by following the steps in “Accessing FCC Manual Control” above.
2. At the D_OUT buttons on the IN OUT screen (see Figure 13), touch 0 and 2. If the pressure is still below 15 psi, de-

energize the devices and verify general troubleshooting steps have been followed before continuing.

Figure 14.   TLS-4xx HydrX Overview Screen

3. If adequate pressure cannot be generated, the STP Adapter ball valve and check valve must be inspected:

Tag, lockout power to the STP.

- Remove and inspect the STP Adapter ball and check valve for any debris that may be restricting flow.
- Verify that the ball valve is opening fully.
- Re-install STP Adapter assembly and repeat Step 2 to check pressure.

Содержание HydrX

Страница 1: ...in this manual Service Technician Certification Previously known as Level 2 3 Contractors holding valid Technician Certifications are approved to perform installation checkout startup programming and...

Страница 2: ...ftware To avoid electric shock do not touch any wiring terminals inside the FCC while performing this procedure 1 If you have a FCC upgrade thumb drive with the software upgrade already on it go to St...

Страница 3: ...w code and user interfaces are downloaded to the FCC Remove the thumb drive and close and secure the FCC cover Section 4 FCC Clock Setup To set the FCC internal clock on initial startup follow the ste...

Страница 4: ...Menu button to return to the System Mode Top Page Screen see Figure 7 then touch the Run button on that page return to the FCC Home screen YYYY MM DD HH MM SS 01 20 01 09 37 9 AC SAVE M a i n M e n u...

Страница 5: ...t filter alarms due to filter loading characteristics Status Comm User programmed custom alarm i e HydrX Service Persistent communications errors between HydrX and ATG Check serial cable connection Ch...

Страница 6: ...are seated properly in the bottom of each filter compartment The opening on the bottom of each filter is keyed to the bottom of each filter compartment You may need to rotate the filter to Status Vacu...

Страница 7: ...nspect the drain valve to ensure it is not stuck open TROUBLESHOOTING TIPS Accessing FCC Manual Control 1 Go to the Self Diag screen on the FCC as follows a Touch and hold the upper left corner of the...

Страница 8: ...1 Enter the FCC Self Diag screen see Figure 11 by following the steps in Accessing FCC Manual Control above 2 At the D_OUT buttons on the IN OUT screen see Figure 13 touch 0 and 2 If the pressure is s...

Страница 9: ...have been followed a If no leaks are visually present and the system cannot hold vacuum during idle periods the following steps can be followed Run a Fill or New Filter cycle Setting Fill Time to 3 mi...

Страница 10: ...e and run a Fill or New Filter cycle If all of the previous steps have been taken contact Veeder Root Technical Support for further instructions Comm Fault 1 The comm parameters for the FCC are 9600 o...

Отзывы: