(mTorr) or millibar. One mTorr is equal to 1 x 10
-3
Torr. One
millibar (mbar) is equal to 1 x 10
-3
bar.
3
To alter the Crossover 1 set point, press either of the “SET”
arrow keys. The alphanumeric display will change either down
or up in correspondence with the number of key presses of the
left or right arrow keys. If a user tries to alter a setting outside
of its allowed range, the piezoelectric beeper within the unit will
emit a short sound. Once the Crossover 1 entry has been
altered, it will remain at the new setting until it is altered once
again. If testing is interrupted by an ERROR 01 or ERROR 05,
Crossover 1 has been set too high for this application and should
be lowered.
Setting Crossover 2
The second “crossover”, Crossover 2, is defined as the set point (the point
during pump down) where the unit switches from reverse mode testing
to direct mode testing. When the unit determines that the leak is smaller than
a “reverse flow” leak, the fine valve opens, the internal rough valve closes
and the unit “crosses over” into direct flow testing.
To reset the Crossover 2 parameter:
1
Press the “SELECT” keys until the green indicator next
to Crossover 2 is lit.
2
The Service Mode alphanumeric display will display a number
between 1 to 500 indicating the crossover set point in millitorr
(mTorr) or millibar. One mTorr is equal to 1 x 10
-3
Torr. One
millibar (mbar) is equal to 1 x 10
-3
bar.
3
To alter the Crossover 2 set point, press either of the “SET”
momentary switches. The alphanumeric display will change
either down or up in correspondence with the number of
key presses of the left or right arrow keys. If a user tries to alter
a setting outside of its allowed range, the piezoelectric beeper
within the unit will emit a short sound. Once the Crossover 2
entry has been altered, it will remain at the new setting until it is
altered once again. If testing is interrupted by an ERROR 02,
Crossover 2 has been set too high for this application and should
be lowered. It may also help to increase the “D1” parameter,
“Fine
Crossover
Delay.”
1109-186-00 Rev. C
2-26
Содержание MS40
Страница 1: ......
Страница 5: ...Figure 1 1 VIC Model MS 40 Portable Automatic Leak Detector 1109 186 00 Rev C 1 2...
Страница 9: ...6 3 4 2 5 7 1 9 8 FIGURE 1 2 INTERNAL ROTARY VANE MECHANICAL PUMP 1109 186 00 Rev C 1 6...
Страница 17: ......
Страница 19: ...FIGURE 2 1 MS 40 USER PANEL 1109 186 00 Rev C 2 2...
Страница 32: ...FIGURE 2 2 MS 40 REAR PANEL CONNECTIONS 1109 186 00 Rev C 2 15...
Страница 35: ...Figure 2 3 MS 40 Hand Held Remote Control 1109 186 00 Rev C 2 18...
Страница 45: ...1109 186 00 Rev C 2 28...
Страница 53: ......
Страница 60: ...1109 186 00 Rev C 3 7...
Страница 69: ......
Страница 85: ...1109 186 00 Rev C 4 16...
Страница 98: ...1109 186 00 Rev C 4 29...
Страница 103: ...1109 186 00 Rev C 4 34...
Страница 104: ...1109 186 00 Rev C 4 35...
Страница 105: ...1109 186 00 Rev C 4 36...
Страница 106: ...1109 186 00 Rev C 4 37...
Страница 107: ...1109 186 00 Rev C 4 38...
Страница 108: ......
Страница 112: ...1109 186 00 Rev C 5 4...
Страница 115: ...1109 186 00 Rev C 5 7...
Страница 118: ......
Страница 122: ......
Страница 136: ...1109 186 00 Rev C b ii...
Страница 137: ...1109 186 00 Rev C b iii...
Страница 138: ......
Страница 160: ......
Страница 162: ......
Страница 164: ...1109 186 00 Rev C f ii...
Страница 165: ...1109 186 00 Rev C f iii...
Страница 167: ...1109 186 00 Rev C f v...
Страница 168: ...1109 186 00 Rev C f vi...
Страница 169: ...1109 186 00 Rev C f vii...