Surface Potential Imaging—Procedure
Rev. D
Dimension 3100 Manual
301
16.6 Surface Potential Imaging—Procedure
1. Locate the two toggle switches on the backside of the Basic Extender box (
),
then verify that they are toggled as shown in
Figure 16.6a
Toggle Switches on Back of Basic Extender Module
Table 16.6a
Basic Extender Module toggle switch settings for surface potential imaging.
Note:
The toggle switch combination of
Surface Potential
=
ON
and
Analog2
=
ON
is
not recommended
and can produce erratic and undefined results.
2. Mount a sample onto the sample holder. Make any external electrical connections that are
necessary for the sample.
3. Mount a metal-coated NanoProbe cantilever into the standard cantilever holder. MFM-style
cantilevers (225µm long, with resonant frequencies around 70kHz) usually work well. It is
also possible to deposit custom coatings on model FESP silicon TappingMode cantilevers.
Mode
Tip or Sample Voltage
FM/Phase
Surface
Potential
GND/Surface
Potential
Analog 2
TappingMode
Contact AFM
MFM
√
√
Surface
Potential
√
√
Apply voltage
to tip or sam-
ple (Use for
electric field
gradient
imaging; tun-
neling AFM)
√
√
FM/Phase
Surface Potential
Gnd/Surface Potential
Analog 2
Mode
Tip or Sample
Voltage
To
NanoScope
To
Microscope
Содержание Dimension 3100
Страница 12: ...xii Dimension 3100 Manual Rev D ...
Страница 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Страница 72: ......
Страница 106: ......
Страница 118: ......
Страница 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Страница 222: ......
Страница 266: ......
Страница 274: ......