background image

  S-19-01-1125-UV101 V1.3 

3. PRODUCT DESCRIPTION & FIELDS OF APPLICATION 

 
The Vector LED-P curing light is a pen-style, cordless, high-performance LED light source curing unit 
intended for polymerization of resin-based light cured dental materials by dental professionals. Vector 
LED-P incorporates 5W blue LED to achieve high energy output with wavelength range of 440 to 480nm 
to polymerize resin-based composite in the market and a high power mode to produce super fast curing 
effect. The relevant range is for camphorquinone (CQ) containing product. Though the majority of 
light-curing dental materials are responsive in this range of wavelengths, it is suggested you should 
contact the filling material maker to make sure it before use. 
 
The Vector LED-P LED curing light is a fast-curing LED curing light. It emits typical light output at about 
1,400 mW/cm². 3 Seconds to cure most resin-based composite in the market. It also features: 
 

► 

Digital timer display, multiple mode selection: boost (fast curing) and ramp up (soft start),

 

► 

Two button easy operation,

 

► 

Quiet operation, fan-free design,

 

► 

Rubber button for more comfortable finger touch, and resist water or solution when cleaning up. 
 

► 

Can be used both cordless and corded, 

► 

Low battery warning; for urgent use, just insert power cord connector into handpiece, 

► 

Rechargeable lithium battery inside,

 capable of 200 x 10 sec curing cycles prior to recharging,

 

► 

Slim and ergonomic handpiece for most comfortable hand holding, (weight 155g only) 
 

► 

8

mm

 diameter

 

sterilizable

 curing tip

,

 

► 

Curing tip rotates 360

for curing at any position of mouth, 

► 

Versatile fiber optic curing tip optional, 

 
 

4. PACKAGE CONTENT 

 

Charging Base Assembly 

X 1 PCE 

Vector LED-P Hand piece Assembly 

X 1 PCE 

8 mm Diameter, Sterilizable Fiber Optic Light Guide 

X 1 PCE 

Protective Eye Shield 

X 1 PCE 

Wall Plug-in Power Supply 

X 1 PCE 

Instructions For Use 

X 1 PCE 

 

Содержание LED-P

Страница 1: ...S 19 01 1125 UV101 V1 3 Vector LED Curing Light Vector LED P Instructions For Use VECTOR R D INC 6824 19TH ST 230 UNIVERSITY PLACE WA 98466 USA Website www vectorusa net User Manual...

Страница 2: ...e Environment 6 Installation of the unit p 7 6 1 Initial Setup 6 2 Keypad And Modes 6 3 Timer Selection 6 4 Battery Charging 7 Operation Instruction p 10 8 Operating Errors Trouble shooting p 11 9 Ser...

Страница 3: ...ON Do not look directly into the light emitted from this curing unit It may be harmful to the eyes Exposure must be restricted to the area of the oral cavity in which clinical treatment is intended Do...

Страница 4: ...Check unit for adequate light output before each procedure Failure to verify output may allow inadequate curing 6 In order to avoid electric shock do not introduce any objects into the unit with the...

Страница 5: ...t output at about 1 400 mW cm 3 Seconds to cure most resin based composite in the market It also features Digital timer display multiple mode selection boost fast curing and ramp up soft start Two but...

Страница 6: ...25 minutes Dimensions Diameter 25 30 mm Length with without light guide 240 mm 180 mm Weight with light guide and battery Roughly 155 g 5 3 Charging Base and Handpiece Time to charge empty battery App...

Страница 7: ...take about 6 hours Use only the power supply furnished with the light system Attempting to use another power supply source may cause injury or damage and will void the warranty 3 Mode button surround...

Страница 8: ...all local requirements Always use unit in conjunction with a properly grounded outlet 6 2 Keypad and modes There are 2 buttons on the key pad for the mode timer selection and the on off button Fig 5...

Страница 9: ...tached to the charging base or attach the power supply directly to the handpiece When this is done the POWER INDICATOR On Off button surrounding will illuminate Figure 7 The light may be used with the...

Страница 10: ...the surface being treated Rotate the light guide into the optimal position for polymerization Place the light guide as close to the filling material as possible Avoid directly contacting the filling m...

Страница 11: ...when placing handpiece on the charging base Check the power plug connects to power outlet socket firmly If not sure what happen contact the dealer 6 Battery working time is significantly shortened Ba...

Страница 12: ...l accessories Part Description Contents LM200 Digital Light Radiometer 200 1x Light meter 200 810 Sterilizable fiber optic light guide 10mm diameter 1x 10mm diameter 809 Sterilizable fiber optic light...

Страница 13: ...moved with alcohol a plastic spatula may help in removing the material Do not use any sharp or pointed tools to protect the surface of the light guide from scratching To disinfect place the cleaned li...

Страница 14: ...re shield with a soft cloth and if required a mild cleaning agent Solvents or abrasive cleaners may not be used in any case as they damage the plastic components Make sure that charge connector remain...

Страница 15: ...t by an authorized engineer dealer warranty will expire immediately 13 3 Warranty Exception The warranty stated herein is the sole warranty applicable to Vector R D products Vector R D expressly discl...

Отзывы: