46
A.3
Error Message
1. USB device driver is not installed.
●
Solution: Please install the device driver again.
2. Control box cannot be initialized
.
●
Solution: Check and re-connect the USB PC interface cable.
3. USB device driver is not working properly.
●
Solution: Re-install the driver.
4. Capture program is already running.
●
Solution: Please close any other programs.
5. Detector response time-out.
●
Check and re-connect the USB PC interface cable. Please try again.
If the same message is displayed again, contact Customer Service.
6. Data communication error.
●
Solution: Re-connect the USB PC interface cable.
7. Canceled image capturing.
●
This means that the user canceled image capture. Please try again.
8. Cannot find dark frame.
●
Solution: Restore the EzSensor
‟s calibration data from the S/W installation CD or re-
calibrate the sensor. If the same message is displayed again, contact Customer Service.
9. Cannot find bright frames for calibration.
●
Solution: Reinstall the EzSensor driver.
10. Bad Pixel Map correction error.
●
Solution: Restore the EzSensor
‟s calibration data from the S/W installation CD or re-
calibrate the sensor. If the same message is displayed again, contact Customer Service.
11. Wrong image processing parameters.
●
Solution: Check the X-ray source. If the problem persists, call for technical assistance.