![Vatech EzScan-i Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/vatech/ezscan-i/ezscan-i_user-manual_3616519090.webp)
Page 90
User Manual
11.9
EzScan Performance
WARNING
: In order to prevent over-heating of the system, the
ventilation opening at the bottom of the handpiece should
never be obstructed.
NOTE
: During operation of the system, the handpiece and the tip
may get slightly warm; this is normal.
WARNING
: Portable RF communication equipment (including
peripherals such as antenna cables and external antennas)
should not be used within 30 cm of any part of the EzScan,
including cables specified by the MANUFACTURER. Otherwise,
performance degradation of this equipment may occur.
WARNING
: Use of this equipment adjacent to or on other
equipment must be avoided as it may result in improper
operation. If this use is necessary, it is advisable that this and
the other equipment be observed to verify that they are
operating normally.
WARNING
: Under normal use, the EzScan should transmit
images to the laptop/notebook with the installed image
manipulation software and the video stream is visible in the
bottom left corner of the screen. If not used properly, there
may be loss of transmission of image information or slow
transmission temporarily.