background image

 

 

Pin             

J8

 








8

 

0 Volts 
Well Common 
Recharge Voltage 
Recharge Voltage 
Recharge Voltage 
Switch Return Well 
Switch Return Well 
Switch Return Well 

Pin            

J9

 









Indirect Feed  
Com: Low Voltage 
Aux2 DC 
Child DC 
Muscle DC 
Fixation DC 
Indirect Return  
6 VAC 
6 VAC 

Pin         

J4

 




Ground 
Neutral 
Hot switch in 
Fused hot return 
(Slit Lamp) 

Pin          

J1                   J2

 










10 
11 
12 

0V 
On 
Off 
Projector 
Muscle 
Fixation 
Aux1 
Lamp 
Dimmer 
Child 
Aux2 
------ 

+5 Volts 
On 
Off 
Projector  
Muscle 
Fixation 
Aux1 
Lamp 
Dimmer 
Pwr LED 
Child 
Aux2 

Pin       

J10

 








Dimmer 
Incandescent 
Dimmer 
Fluorescent 
Projector 
Lamp 
Aux1 
Slit Lamp 
Return 
 

F1 = 6 ¼ ASB: lamp, slit lamp, & auxiliary 
F2 = 6 ¼ ASB: room lights 
F3 = ¾ ASB: transformer primary 
F4 = 4 ASB: transformer secondary 

 

J3 Jumper

: For fluorescent and indirect. 

J6 Jumper:

 For incandescent light only. 

Содержание Z9000 Series

Страница 1: ...620 Parkway Broomall PA 19008 610 356 3995 www varitronics com Varitronics Inc Z9000 Series Manual Copyright 2002 Varitronics Inc...

Страница 2: ...rly connected Take the proper steps to turn off the power leaving all personnel and patients in safe condition if anything is found wrong with the system See Appendix C Be sure to observe the followin...

Страница 3: ...he safest systems on the market today It has been engineered with many safety features to prevent shock and fire hazard In addition there are no electromagnetic interference concerns Be sure to observ...

Страница 4: ...5 ADJUSTMENTS Programming the Z9000 Ophthalmic Control System 6 MAINTENECE 7 TROUBLESHOOTING 8 TERMS OF TRANSPORTATION APPENDIX A Z9000 PC Board Schematic APPENDIX B Z9000 Wiring Diagrams APPENDIX C...

Страница 5: ...has configurable jumpers J3 J6 to provide incandescent or incandescent with fluorescent room lighting control See Appendix A for jumper settings 1 2 Termination Box A 1 2 1 Termination Box A requires...

Страница 6: ...wer box will vary depending on which model is purchased These units are meant to work with 115 VAC only 2 1 4 Have the doctor or architect check all intended locations before installing ATTENTION Inst...

Страница 7: ...d neutral voltage lines The fused circuit must be on the hot voltage line Please see electrical schematics for exact details Determine the location for Box B Box B the connections for low voltage func...

Страница 8: ...ixtures These lamps are to be wired in parallel Wires shall run to Power Box A wire 7 Maximum load 300 watts 3 9 FINAL NOTES 3 9 1 For final installation please refer to Varitronics wiring schematic 3...

Страница 9: ...the indirect is lifted 4 6 When the slit lamp control switch optional is turned on it will supply power to the slit lamp and dim the room lights or turn them off Maximum load 60 watts 4 7 The Z9000 Op...

Страница 10: ...ndirect Ophthalmoscope Room lights dim 2 Indirect Ophthalmoscope Room lights off 3 Projector Room lights dim default to on 4 Not used 5 Not used 6 Not used 6 MAINTINENCE 6 1 There are no periodic main...

Страница 11: ...ll only change the three external fuses A qualified technician should change any internal fuses See Appendix A 6 3 1 First turn the power switch off and disconnect the main AC plug on the power supply...

Страница 12: ...battery date code If beyond date replace batteries 7 1 5 Problem Auxiliary Lamp Slit Lamp not working Solution Replace Fuse marked F1 Refer to Appendix A 7 1 6 Problem System is completely non functio...

Страница 13: ...APPENDIX A...

Страница 14: ...switch in Fused hot return Slit Lamp Pin J1 J2 1 2 3 4 5 6 7 8 9 10 11 12 0V On Off Projector Muscle Fixation Aux1 Lamp Dimmer Child Aux2 5 Volts On Off Projector Muscle Fixation Aux1 Lamp Dimmer Pwr...

Страница 15: ...APPENDIX B...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ...APPENDIX C...

Страница 20: ...ASB 4 ASB INTERIOR LABELS WARNING For continued protection against fire hazard replace with same type and rating of fuse only MODEL Z9000 Desktop 1P 2P Full Load 700 Watts 115 VAC 60 Hz Only Varitron...

Страница 21: ...APPENDIX D...

Страница 22: ...Specification Sheet Z9000...

Страница 23: ...Specification Sheet Z9000 1P...

Страница 24: ...Specification Sheet Z9001 2P...

Страница 25: ...Specification Sheet FW9001 2P...

Страница 26: ...Specification Sheet Z9001 2P...

Страница 27: ...Specification Sheet Z9001 2P...

Страница 28: ...APPENDIX E...

Страница 29: ...PHOTOS OF Z9000 SERIES OPHTHALMIC CONTROL SYSTEMS...

Страница 30: ...PHOTOS OF ACCESSORIES...

Страница 31: ...PHOTOS OF ACCESSORIES...

Отзывы: