background image

 

 

5.1.1 

Switch Bank 1- First Well Dipswitch 

 

 

 

Dipswitch Function 

Muscle light

 

Fixation 

 

Room lights dim

 

Room lights off

 

 

 

 

  

5.1.2

 Switch Bank 1- Second Well Dipswitch 

 

 

 

 

 

 

 

Dipswitch Function 

Muscle light

 

Fixation 

 

Room lights dim

 

Room lights off

 

 
 

 

5.1.3

 Switch Bank 1- Third Well Dipswitch 

 

 

 

 

Dipswitch Function 

 9 

Muscle light

 

10 

Fixation 

 

11 

Room lights dim

 

12 

Room lights off

 

 

 

 

5.1.4

 Switch Bank 2 

 

 

 

  

 

 

 

Dipswitch Function 

Indirect Ophthalmoscope- 
Room lights dim

 

Indirect Ophthalmoscope- 
Room lights off

 

Projector- Room lights dim 
(default to on) 

4 Not 

used 

      Not used 

      Not used

 

 
 

6. MAINTINENCE 

 

 

6.1

 There are no periodic maintenance requirements for the Z9000. Bulbs and external 

 

fuses should be replaced on an ‘as needed’ basis.  Refer to section 6.2 for the proper 

 

replacement method for bulbs and section 

6.3 

for external fuses. 

 

 

6.2 

Testing and replacing lamp bulbs. If the lamp does not light, check to make sure that   the 

 

associated switch is activated. If the lamp still does not light, replace the bulb with the 

 

same kind as instructed below. 

Содержание Z9000 Series

Страница 1: ...620 Parkway Broomall PA 19008 610 356 3995 www varitronics com Varitronics Inc Z9000 Series Manual Copyright 2002 Varitronics Inc...

Страница 2: ...rly connected Take the proper steps to turn off the power leaving all personnel and patients in safe condition if anything is found wrong with the system See Appendix C Be sure to observe the followin...

Страница 3: ...he safest systems on the market today It has been engineered with many safety features to prevent shock and fire hazard In addition there are no electromagnetic interference concerns Be sure to observ...

Страница 4: ...5 ADJUSTMENTS Programming the Z9000 Ophthalmic Control System 6 MAINTENECE 7 TROUBLESHOOTING 8 TERMS OF TRANSPORTATION APPENDIX A Z9000 PC Board Schematic APPENDIX B Z9000 Wiring Diagrams APPENDIX C...

Страница 5: ...has configurable jumpers J3 J6 to provide incandescent or incandescent with fluorescent room lighting control See Appendix A for jumper settings 1 2 Termination Box A 1 2 1 Termination Box A requires...

Страница 6: ...wer box will vary depending on which model is purchased These units are meant to work with 115 VAC only 2 1 4 Have the doctor or architect check all intended locations before installing ATTENTION Inst...

Страница 7: ...d neutral voltage lines The fused circuit must be on the hot voltage line Please see electrical schematics for exact details Determine the location for Box B Box B the connections for low voltage func...

Страница 8: ...ixtures These lamps are to be wired in parallel Wires shall run to Power Box A wire 7 Maximum load 300 watts 3 9 FINAL NOTES 3 9 1 For final installation please refer to Varitronics wiring schematic 3...

Страница 9: ...the indirect is lifted 4 6 When the slit lamp control switch optional is turned on it will supply power to the slit lamp and dim the room lights or turn them off Maximum load 60 watts 4 7 The Z9000 Op...

Страница 10: ...ndirect Ophthalmoscope Room lights dim 2 Indirect Ophthalmoscope Room lights off 3 Projector Room lights dim default to on 4 Not used 5 Not used 6 Not used 6 MAINTINENCE 6 1 There are no periodic main...

Страница 11: ...ll only change the three external fuses A qualified technician should change any internal fuses See Appendix A 6 3 1 First turn the power switch off and disconnect the main AC plug on the power supply...

Страница 12: ...battery date code If beyond date replace batteries 7 1 5 Problem Auxiliary Lamp Slit Lamp not working Solution Replace Fuse marked F1 Refer to Appendix A 7 1 6 Problem System is completely non functio...

Страница 13: ...APPENDIX A...

Страница 14: ...switch in Fused hot return Slit Lamp Pin J1 J2 1 2 3 4 5 6 7 8 9 10 11 12 0V On Off Projector Muscle Fixation Aux1 Lamp Dimmer Child Aux2 5 Volts On Off Projector Muscle Fixation Aux1 Lamp Dimmer Pwr...

Страница 15: ...APPENDIX B...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ...APPENDIX C...

Страница 20: ...ASB 4 ASB INTERIOR LABELS WARNING For continued protection against fire hazard replace with same type and rating of fuse only MODEL Z9000 Desktop 1P 2P Full Load 700 Watts 115 VAC 60 Hz Only Varitron...

Страница 21: ...APPENDIX D...

Страница 22: ...Specification Sheet Z9000...

Страница 23: ...Specification Sheet Z9000 1P...

Страница 24: ...Specification Sheet Z9001 2P...

Страница 25: ...Specification Sheet FW9001 2P...

Страница 26: ...Specification Sheet Z9001 2P...

Страница 27: ...Specification Sheet Z9001 2P...

Страница 28: ...APPENDIX E...

Страница 29: ...PHOTOS OF Z9000 SERIES OPHTHALMIC CONTROL SYSTEMS...

Страница 30: ...PHOTOS OF ACCESSORIES...

Страница 31: ...PHOTOS OF ACCESSORIES...