![Varian TV 301 Navigator 969-8918 Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/varian/tv-301-navigator-969-8918/tv-301-navigator-969-8918_instruction-manual_3709206048.webp)
TECHNICAL INFORMATION
TYPICAL LAYOUT DIAGRAM
Side mounting
NOTE
With Navigator Controller
The L-shaped bracket (P/N 969-9349) is available
as an option.
See the following figure.
1. Screw the 4 fixing studs (provided with the ac-
cessories bag) in the holes on the L-shaped
bracket;
2. Place the controller on the studs aligning the 4
rubber holes on the top of the controller;
3. Press the controller towards the bracket;
4. Turn the pump upside-down;
5. Place the bracket on the pump bottom;
6. Screw a little bit the three screws M4 in the
holes on the pump bottom;
7. Rotate the bracket until the controller is on the
chosen position;
8. Tighten the screws properly;
1. Turbo-V Navigator controller
9. Turn the pump again;
2. Vent
valve
10. Plug the line cord and connect the pump cable
to the pump;
3. Vacuum pump shut-off valve (optional)
4. System vent valve (optional)
11. Connect the mating connector with the jumper
on the interlock signal to start the pump.
5. Vacuum
chamber
MOUNTING 2
6. Ionization
gauge
7. Fore-vacuum pump connecting flange
8. Oil mist eliminator
9. Fore-vacuum pump with internal one-way
valve
10. Fore-vacuum pump control relay
11. Connection for water cooling
12. Roughing line with valve (optional)
13. Turbopump
14. Fan
15. Flexible connection
43