TECHNICAL INFORMATION
N. Read/
Write
Data
Type
Description Admitted
Values
300 R N Cycle
time
in minutes
(zeroed by
the reset
command)
0 to 4x10
9
301 R N Cycle
num-
ber (zeroed
by the reset
command)
0 to 4x10
9
302
R
N
Pump life in
hours (ze-
roed by the
reset com-
mand)
0 to 4x10
9
320
to
399
Reserved to Varian service
400 R A CRC
EPROM
(QE)
QE5XXXX
(where “XXXX”
are variable)
402 R A CRC
Pa-
ram. (PA)
PA5XXXX
(where “XXXX”
are variable)
500
Reserved to Varian service
503
R/W
N
RS 485 ad-
dress
0 to 31
(default = 0)
504 R/W L Serial
type
select
0 = RS 232
1 = RS 485
(default = 0)
NOTES
1. Automatic means that when the controller
stops, the vent valve is opened with a delay de-
fined by window n. 126; when the controller
starts, the vent valve is immediately closed.
2. On command means that the vent valve is
opened or closed by means of window n. 122.
3. These commands remove the pump motor
supply for a short time. If they are used con-
tinuously, they can cause the pump stopping.
7 6 5 4 3 2 1 0
TOO
HIGH
LOAD
NO
CONNECTION
SHORT
CIRCUIT
PUMP
OVERTEMP.
OVERVOLTAGE CONTROLL.
OVERTEMP.
AUX FAIL POWER FAIL
Window N. 206 Bit Description
INLET SCREEN INSTALLATION
The inlet screen mod. 969-9316 prevents the
blades of the pump from being damaged by debris
greater than 0.7 mm diameter.
The inlet screen, however, does reduce the pump-
ing speed by about 10%.
The inlet screen is fitted in the upper part of the
pump, as shown in the figure.
The screen can be mounted on each pump.
The screen can be removed as shown in the fol-
lowing figure.
35