TECHNICAL INFORMATION
Controller Description
The dedicated controller is a solid-state frequency
converter which is driven by a single chip micro-
computer and is composed of two PCBs which in-
clude power supply and 3-phase output, analog
and input/output section, microprocessor and digi-
tal section.
The power supply, together with the 3-phase out-
put, converts the single phase AC mains supply
into a 3-phase, low voltage, medium frequency
output which is required to power the pump.
The controller can be operated by a remote host
computer via the serial connection. A Windows-
based software is available (optional).
TECHNICAL SPECIFICATION
Pumping speed (with
inlet screen)
N
2
:
He:
H
2
:
ISO 200K / ISO 200F:
950 l/s
870 l/s
900 l/s
Compression ratio
N
2
He
H
2
1 x 10
9
1 x 10
7
1 x 10
6
Base pressure
with recommended
forepump
<
1 x 10
-10
mbar
(
<
1 x 10
-10
Torr)
(
According to standard DIN 28 428,
the base pressure is that measured
in a leak-free test dome, 48 hours
after the completion of test dome
bake-out, with a Turbopump fitted
with a ConFlat flange and using the
recommended pre-vacuum pump)
Inlet flange
ISO 200K, ISO 200F
Foreline flange
KF 40 NW
Rotational speed
38000 rpm
Start-up time
<
4 minutes
Recommended
forepump
DS 402, TriScroll 300
Operating position
Any
Operating ambient
temperature
+5 °C to +35 °C
Bakeout temperature
80° C at inlet flange max.
Vibration level
(Data measured with
an indicated bearing
temperature of 20
°C: see vibration
graph in the following
figure)
bearing peak: < 45 mm/sec
2
residual unbalance: < 350 mm/sec
2
(see the following typical vibration
graph)
Lubricant permanent
lubrication
Cooling require-
ments
Forced air or water
Coolant water
Minimum flow: 200 l/h (0.89 GPM)
Temperature: +10 °C to +20 °C
Pressure: 3 to 5 bar (45 to 75 psi)
Noise level
<45 dB(A) at 1 meter
Power supply:
Input
voltage:
Input
freq.:
Max input power:
Stand-by
power:
Max
operating
power:
100 - 240 Vac
50 - 60 Hz
850 VA
30 to 35 W
450 W with water cooling
220 W with air cooling
Protection fuse
1 x 6.3 A
Compliance with:
UNI EN 292-1
UNI EN 292-2
EN-CENELEC 55011
IEC 1000-4-2 (ex 801-2)
IEC 1000-4-3 (ex 801-3)
IEC 1000-4-4 (ex 801-4)
EN 61010-1 (IEC 1010-1)
EN 1012-2
Installation category
II
Pollution degree
2
Power cable
With European or NEMA plug
3 meters long (optional)
Serial communica-
tion (Navigator kit)
RS232 cable with a 9-pin D type
male connector and a 9-pin D type
female connector, and Navigator
software (optional)
Storage temperature
-20° C to +70° C
Weight kg (lbs)
ISO 200K flange
ISO 200F flange
19 (41.8)
19.4 (43)
Controller
5.4
(12)
NOTE
When the TV 1001 SEM has been stored at a
temperature less than 5°C, wait until the system has
reached the above mentioned temperature.
26
87-900-973-01