TECHNICAL INFORMATION
7
1 87-900-870-01 (
E
)
Inspect the controller for any shipping damage.
!
High voltage developed in the controller can cause
severe injury or death. Before servicing the unit,
disconnect the input power cable.
NOTE
The Turbo-V controller must be positioned so that free
air can flow through the holes. Connection to the mains
must be made in accordance with the local law. Always
connect the ground wire and use a properly grounded
power socket to avoid electrical shock.
Line Voltage Change Over
The controller is factory-set for 120 Vac operation.
If a change in line voltage operation is desired, proceed
as follows:
−
Disconnect the INPUT/OUTPUT connector J17.
−
Select the desired operating voltage on left hand
panel.
−
Check voltage selector window for correct set and
insert connector J17.
Input/Output Connector
Input/output signals and mains must be connected to
J17 mating connector (not provided). Make the
connections with AWG 20 (0.5 mm
2
) to the pins
indicated in the following figure, to obtain the desired
capability.
P17 I/O Connector