It is possible to delay the set point checking for a
programmable delay time.
The PROGRAMMABLE SET POINT signal has the
following default settings:
−
reference quantity: frequency
−
thereshold: 867 Hz
−
hysteresis:
2%
−
activation type: high level
−
delay time: 0 second
NOTE
The Navigator Software (optional) allows the op-
erator to set all the programmable feature.
How to connect the open collector input of the
controller
Here below there are the typical connections of the
open collector input of TPS-compact to an external
system. Two cases are considered:
1.
the customer supplies the 24 Vdc
2.
the customer does not supply the 24 Vdc.
Please note that on the connector a 24 Vdc, 30
mA voltage, a GROUND signal and the open col-
lector pin are available.
Case 1
Case 2 with relay utilisation
Case 2 with transistor utilisation