ProStar 400 Autosampler Control Manual
-32-
03-914947-28:R1
The temperature setpoint can be changed. Enter the new
temperature setpoint and then click on the
button to send the new value to the autosampler.
Click on the
button to start the sample tray
cooling.
Click on the
button to stop the sample tray
cooling.
The
Temperature circuit error
LED indicates if an error has
occurred.
NOTE :
Although it is possible to program setpoints from 0 °C to 40 °C, the ProStar 400
Autosampler can only control the temperature down to ambient tempe5°C.
NOTE :
The temperature range for sample tray cooling is from 4 °C to 15 °C and the
maximum cooling capacity is 20 °C below ambient temperature.
Overview Window
Click on the
button: the
Status Overview
window contains
three tabs:
•
General:
displays the synopsis of the instrument,
where some current values can be read (vial
number, volume, etc.).