22
03-914783-00:3
Figure 9 Flowcell and Lamp Assemblies
To change the flowcell:
WARNING:
SHOCK HAZARD
Make sure the power cord is disconnected from the rear panel of the detector.
1. Disconnect the cell inlet tube from the column or connecting
tube and free the cell outlet tubing.
2. Remove the cell cover by unscrewing the thumbscrew and
pulling the cover straight back to expose the photodiode
mount.
3. Unscrew the two thumbscrews on the photodiode mount and
pull the photodiode mount straight back (see figure above).
The connecting cable is long enough to allow the photodiode
mount to rest on the bench top.
1.
Cell Cover
2.
Photodiode Retaining
Screw
3.
Photodiode Mount
4.
Tubing Strain Relief
Thrumbscrew
5.
Cell Retaining Screws
6.
Tubing Strain Relief
Clamp
7.
Cell Body Assembly
8.
Cell Holder
9.
Cell holder Tumbscrews
10. Lamp Light Shield
Screw
11. Lamp Light Shield
12. Deuterium Lamp
Assembly and Plug
13. Lamp Thrumbscrew
Standoff
14. Lamp Thrumbscrew
15. Lamp Cover
16. Lamp Cover Screw