ProStar 335 Photodiode Array Detector Control Manual - 24 -
Building a Method
Creating a New Method
The method contains parameters for instrument control, data
acquisition, chromatogram processing, and editing of final
results. To initially set up a method, it is not necessary to define
all of the sub-method sections, however, you must fill in the
instrument control section prior to starting an acquisition.
1. To create a new method, select the FILE / NEW METHOD
option from the Galaxie Chromatography Data System main
menu.
A wizard guide will appear and provide assistance during
the first steps of method creation: